Related Experiment Video
Updated: Aug 9, 2025

Author Spotlight: Advancing Facial Rejuvenation Therapy with Post-Laser Salicylic Acid Application
Published on: September 27, 2024
Safety and Efficacy of a Thermomechanical Fractional Injury Device for Periorbital Rhytides
Jordan V Wang1, Shirin Bajaj1, David Orbuch1
1Laser & Skin Surgery Center of New York, New York, New York.
Background:
Periorbital rejuvenation is a common cosmetic concern. A fractional thermomechanical skin rejuvenation system was developed to offer clinical improvements from direct heat transfer.
Objective:
A prospective study evaluated the efficacy and safety of the device for periorbital fine lines and wrinkles.
Materials And Methods:
Subjects with moderate-to-severe periorbital rhytides were enrolled and underwent 4 monthly treatments with a novel device using thermomechanical fractional injury (Tixel 2; Novoxel, Netanya, Israel).
Results:
Fifty-one subjects were enrolled. Mean age was 56.8 years, and 88.2% were women. Fitzpatrick skin Types I to IV were included. For Fitzpatrick Wrinkle Classification System (FWCS), mean baseline score was 5.7. Per investigator, there was a mean 2.0-grade improvement in FWCS at 3-month follow-up ( p < .0001). Per 3 blinded physician raters, there was a mean improvement of 2.2 ( p < .0001), 2.0 ( p < .0001), and 1.2 ( p < .0001) in FWCS at 3-month follow-up. Each of the raters correctly identified posttreatment images for 87.5%, 77.1%, and 75.0% of subjects. At least 2 raters agreed on grading 83.3% of subjects as responders. There were no severe adverse events. Subjects experienced minimal pain and downtime.
Conclusion:
A novel device using thermomechanical fractional injury was demonstrated to be safe and effective in the treatment of periorbital rhytides.
More Related Videos
04:20A Saline/Bipolar Radiofrequency Energy Device As an Adjunct for Hemostasis in Solid Organ Injury/Trauma
Published on: July 28, 2020
11:46Investigation of the Electrophysiological and Thermographic Safety Parameters of Surgical Energy Devices During Thyroid and Parathyroid Surgery in a Porcine Model
Published on: October 13, 2022