Related Experiment Video
Updated: Aug 9, 2025

The Rigid Tube as an Alternative in Controlling the Problematic Airway
Published on: June 6, 2020
Anti-Toothbreaker: A Novel Low-Budget Device Enabling Contactless Dental Protection and a Forbidden Technique during
Sam Razaeian1, Helena Kristin Liebich2
1Department of Trauma Surgery, Hannover Medical School, Carl-Neuberg-Straße 1, 30625 Hannover, Germany.
Background:
Iatrogenic dental injury is the most common complication of conventional laryngoscopy during orotracheal intubation. The main cause is unintended pressure and leverage forces from the hard metal blade of the laryngoscope. The aim of this pilot study was to introduce and test a novel, reusable low-budget device not only providing contactless dental protection during direct laryngoscopy for endotracheal intubation, but also enabling, in contrast to established tooth protectors, active levering with conventional laryngoscopes for easier visualization of the glottis.
Methods:
A constructed prototype for intrahospital usage was evaluated by seven participants on a simulation manikin for airway management. Endotracheal intubation was performed with and without the device using a conventional Macintosh laryngoscope (blade size 4) and a 7.5 mm endotracheal tube (Teleflex Medical GmbH, Fellbach, Germany). Necessary time and success of first pass were determined. Degree of visualization of the glottis with and without the device was stated by the participants according to the Cormack and Lehane (CL) classification system and the Percentage of Glottic Opening (POGO) scoring system. In addition, subjective physical effort, feeling of safety regarding successful intubation, and risk for dental injury were queried on a numeric scale between 1 and 10.
Results:
All participants except one stated that the intubation procedure was easier with usage of the device than without it. On average, this was subjectively perceived as being approximately 42% (range, 15-65%) easier. In addition, time to first pass success, as well as degree of glottis visualization, subjective physical effort, and feeling of safety regarding risk for dental injury, were clearly better with usage of the device. Concerning feeling of safety regarding successful intubation, there was only a minor advantage. No difference in first pass success rate and number of total attempts could be observed.
Conclusion:
The Anti-Toothbreaker is a novel, reusable low-budget device which might not only provide contactless dental protection during direct laryngoscopy for endotracheal intubation, but also enables, in contrast to established tooth protectors, active levering with conventional laryngoscopes for easier visualization of the glottis. Future human cadaveric studies are needed to investigate whether these advantages also prove themselves there.
Related Concept Videos
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Tracheostomy Care I: Pre-procedural Steps
Required Equipment
The equipment necessary for tracheostomy care includes:
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...