Efficient SARS-CoV-2 Quantitative Reverse Transcriptase PCR Saliva Diagnostic Strategy utilizing Open-Source Pipetting Robots
Published on: February 11, 2022
Accurate Detection of SARS-CoV-2 by Next-Generation Sequencing in Low Viral Load Specimens
Marius Ilié1,2,3, Jonathan Benzaquen3,4, Véronique Hofman1,2,3
1Laboratory of Clinical and Experimental Pathology, Centre Hospitalier Universitaire de Nice, FHU OncoAge, Université Côte d'Azur, 06000 Nice, France.
Abstract:
As new SARS-CoV-2 variants emerge, there is an urgent need to increase the efficiency and availability of viral genome sequencing, notably to detect the lineage in samples with a low viral load. SARS-CoV-2 genome next-generation sequencing (NGS) was performed retrospectively in a single center on 175 positive samples from individuals. An automated workflow used the Ion AmpliSeq SARS-CoV-2 Insight Research Assay on the Genexus Sequencer. All samples were collected in the metropolitan area of the city of Nice (France) over a period of 32 weeks (from 19 July 2021 to 11 February 2022). In total, 76% of cases were identified with a low viral load (Ct ≥ 32, and ≤200 copies/µL). The NGS analysis was successful in 91% of cases, among which 57% of cases harbored the Delta variant, and 34% the Omicron BA.1.1 variant. Only 9% of cases had unreadable sequences. There was no significant difference in the viral load in patients infected with the Omicron variant compared to the Delta variant (Ct values, p = 0.0507; copy number, p = 0.252). We show that the NGS analysis of the SARS-CoV-2 genome provides reliable detection of the Delta and Omicron SARS-CoV-2 variants in low viral load samples.