Metallic Artifact Reduction in Midfacial CT Scans Using Patient-Specific Polymer Implants Enhances Image Quality
Julian Lommen1, Lara Schorn1, Christoph Sproll1
1Department of Oral, Maxillofacial and Facial Plastic Surgery, University Hospital Düsseldorf, Moorenstraße 5, 40225 Düsseldorf, Germany.
Polymer implants for midfacial reconstruction significantly reduce CT imaging artifacts compared to titanium. This improves image quality, aiding in postoperative radiation therapy planning and tumor aftercare.
Area of Science:
- Medical Imaging
- Biomaterials Science
- Oral and Maxillofacial Surgery
Background:
- Midfacial reconstruction often uses bone grafts or implants.
- Titanium implants are common but cause CT imaging artifacts.
- These artifacts complicate treatment planning and follow-up.
Purpose of the Study:
- To evaluate polymer implants for reducing CT metallic artifacts.
- To compare image quality between titanium and polymer midfacial implants.
- To assess the impact on diagnostic imaging for tumor resection cases.
Main Methods:
- Implanted titanium (n=1) and polymer (n=12) zygomatic implants in a human skull specimen.
- Analyzed streak and blooming artifacts using Hounsfield Unit values.
- Assessed overall image quality and compared materials using statistical tests.
Main Results:
- Titanium and hydroxyapatite polymers showed significantly more streak artifacts.
- No significant difference in blooming artifacts was observed between materials.
- Polymer implants resulted in slightly better image quality than titanium.
Conclusions:
- Personalized polymer implants effectively reduce metallic artifacts in CT imaging.
- Improved image quality facilitates postoperative radiation therapy planning.
- Polymer implants enhance radiological tumor aftercare around midfacial reconstructions.