Validation of a blood gas device for ionized calcium analysis in Holstein cows

Abdülkerim Deniz1, Mert Metin2, Kemal Aksoy3

  • 1Free Researcher for Clinical Biochemistry, İstanbul, Turkey.

Insights

The Edan i15 Vet device showed good accuracy for diagnosing subclinical hypocalcemia in dairy cows, though it cannot replace the Gem Premier 3000 analyzer. Serum total calcium measurements were less effective for diagnosis.

Area of Science:

  • Veterinary Medicine
  • Animal Health
  • Dairy Cow Management

Background:

  • Accurate ionized calcium (iCa) analysis is crucial for detecting hypocalcemia in dairy cows.
  • The Edan i15 Vet (EDAN) device's performance for iCa in dairy cows requires validation.

Purpose of the Study:

  • Validate the EDAN blood gas device against the Gem Premier 3000 (GEM) analyzer for iCa measurements.
  • Evaluate the diagnostic capability of EDAN iCa and serum total calcium (TCa) for subclinical hypocalcemia.

Main Methods:

  • Measured iCa using EDAN and GEM devices in 125 lactating Holstein cows (post-calving to day 27).
  • Measured serum TCa using a wet biochemistry method.
  • Analyzed agreement and diagnostic performance (sensitivity/specificity) using Bland-Altman and Passing-Bablok regressions.

Main Results:

  • EDAN device showed a mean bias of 0.05 mmol/L against GEM.
  • EDAN device achieved satisfactory sensitivity/specificity (e.g., 93%/94% at <1.00 mmol/L iCa cutoff) compared to GEM.
  • Serum TCa measurements demonstrated lower sensitivity/specificity for diagnosing subclinical hypocalcemia compared to iCa.

Conclusions:

  • The EDAN device is not interchangeable with GEM but has adequate diagnostic performance for subclinical hypocalcemia based on GEM-derived iCa cutoffs.
  • Serum TCa cutoffs are not suitable for diagnosing subclinical hypocalcemia due to insufficient sensitivity/specificity.
  • iCa values from both devices align with existing literature.
Abstract