Comparison of 50° handheld fundus camera versus ultra-widefield table-top fundus camera for diabetic retinopathy
Edoardo Midena1,2, Luca Zennaro3, Cristian Lapo3
1Department of Ophthalmology, University of Padova, Padova, Italy. edoardo.midena@unipd.it.
Objectives:
To compare the performance of a handheld fundus camera with standard 50° visual field to ultra-widefield (UWF) table-top fundus camera in diabetic retinopathy (DR) detection and grading.
Methods:
Patients affected by diabetes mellitus and referred to our diabetic retinopathy clinic were enroled and underwent fundus photography in mydriasis. All photos were taken using the ultra-widefield table-top fundus camera Zeiss Clarus™ 500 (four fields per eye) and the Optomed Aurora® handheld fundus camera (3 fields per eye). The following parameters were analysed: the gradability of the images, the grade of DR, and diabetic maculopathy (DM), the presence of hypertensive retinopathy (HR) and the presence of other ocular diseases.
Results:
We enroled 759 eyes of 384 diabetic patients and analysed 5313 fundus photos. The handheld fundus camera obtained a sensitivity of 84.2% and specificity of 95.4% for referable cases. Moreover, it obtained, compared to UWF, an almost perfect agreement with linear weighting for DR, DM and HR (k = 0.877, k = 0.854, and k = 0.961, respectively). The lowest sensitivity was achieved for proliferative DR (58.7% sensitivity, 100% specificity).
Conclusions:
Optomed Aurora® handheld fundus camera imaging showed a strong agreement compared to UWF in grading DR, considering all DR and DM grades, in mydriasis. However, the use of UWF imaging increases the detection of referable eyes.