Related Experiment Video
Updated: Aug 8, 2025

3D Planning and Printing of Patient Specific Implants for Reconstruction of Bony Defects
Published on: August 4, 2020
Patient-specific miniplates versus patient-specific reconstruction plate: A biomechanical comparison with 3D-printed
Claudius Steffen1, Kay Sellenschloh2, Magdalena Willsch2
1Charité - Universitätsmedizin Berlin, Corporate Member of Freie Universität Berlin and Humboldt Universität zu Berlin, Department of Oral and Maxillofacial Surgery, Augustenburger Platz 1, 13353, Berlin, Germany.
Background:
Patient-specific 3D-printed miniplates for free flap fixation in mandibular reconstruction were recently associated with enhanced osseous union. Higher mechanical strains resulting from these plates are discussed as reasons, but biomechanical studies are missing. This study aims to examine, whether patient-specific 3D-printed miniplates provide an increased interosteotomy movement (IOM) and lower stiffness compared with reconstruction plates.
Methods:
Polyurethane (PU) mandible and fibula models (Synbone AG, Malans, Schweiz) were used to simulate mandibular reconstruction with a one segment fibula flap equivalent. Osteosynthesis was performed using either four patient-specific 3D-printed miniplates (3D-Mini) or one patient-specific 3D-printed reconstruction plate (3D-Recon). Mastication was simulated using cyclic dynamic loading with increasing loads until material failure or a maximum load of 1000 N. Continuous IOM recording was carried out using a 3D optical tracking system (ARAMIS, Carl Zeiss GOM Metrology, Braunschweig, Germany).
Findings:
The averaged stiffness at a load of 100-300 N load did not differ between the groups (p = 0.296). There was a faster 1.0 mm vertical displacement in the 3D-Mini group (26 376 ± 14 190 cycles versus 44 817 ± 30 430 cycles, p = 0.018). The IOM were higher with miniplate fixation in the distal gap (p = 0.040). In the mesial gap, there was no significant difference between the groups (p = 0.160).
Interpretation:
Fixation with patient-specific 3D-printed miniplates results in higher mechanical strains. Lower rates of pseudarthrosis, as seen in clinical studies, might be caused by this phenomenon. Surgeons should evaluate the primary use of 3D-printed miniplates in mandibular reconstruction due to advantages of intraoral plate removal alongside safe osteosynthesis.