Development and clinical application of a GPU-based Monte Carlo dose verification module and software for 1.5 T
Bo Cheng1, Yuan Xu2, Shijun Li1
1School of Nuclear Science and Technology, University of Science and Technology of China, Hefei, China.
Background:
Adaptive radiotherapy (ART) has made significant advances owing to magnetic resonance linear accelerator (MR-LINAC), which provides superior soft-tissue contrast, fast speed and rich functional magnetic resonance imaging (MRI) to guide radiotherapy. Independent dose verification plays a critical role in discovering errors, while several challenges remain in MR-LINAC.
Purpose:
A Monte Carlo-based GPU-accelerated dose verification module for Unity is proposed and integrated into the commercial software ArcherQA to achieve fast and accurate quality assurance (QA) for online ART.
Methods:
Electron or positron motion in a magnetic field was implemented, and a material-dependent step-length limit method was used to trade off speed and accuracy. Transport was verified by dose comparison with EGSnrc in three A-B-A phantoms. Then, an accurate Monte Carlo-based Unity machine model was built in ArcherQA, including an MR-LINAC head, cryostat, coils, and treatment couch. In particular, a mixed model combining measured attenuation and homogeneous geometry was adopted for the cryostat. Several parameters in the LINAC model were tuned to commission it in the water tank. An alternating open-closed MLC plan on solid water measured with EBT-XD film was used to verify the LINAC model. Finally, the ArcherQA dose was compared with ArcCHECK measurements and GPUMCD in 30 clinical cases through the gamma test.
Results:
ArcherQA and EGSnrc were well matched in three A-B-A phantom tests, and the relative dose difference (RDD) was less than 1.6% in the homogenous region. A Unity model was commissioned in the water tank, and the RDD in the homogenous region was less than 2%. In the alternating open-closed MLC plan, the gamma result (3%/3 mm) between ArcherQA and Film was 96.55%, better than the gamma result between GPUMCD and Film (92.13%). In 30 clinical cases, the mean three-dimensional (3D) gamma result (3%/2 mm) was 99.36% ± 1.28% between ArcherQA and ArcCHECK for the QA plans and 99.27% ± 1.04% between ArcherQA and GPUMCD for the clinical patient plans. The average dose calculation time was 106 s in all clinical patient plans.
Conclusions:
A GPU-accelerated Monte Carlo-based dose verification module was developed and built for the Unity MR-LINAC. The fast speed and high accuracy were proven by comparison with EGSnrc, commission data, the ArcCHECK measurement dose, and the GPUMCD dose. This module can achieve fast and accurate independent dose verification for Unity.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiation: Applications
The average...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...