Related Experiment Video
Updated: Aug 8, 2025

Multi-analyte Biochip MAB Based on All-solid-state Ion-selective Electrodes ASSISE for Physiological Research
Published on: April 18, 2013
All-solid-state potentiometric salicylic acid sensor for in-situ measurement of plant
Cheng-Xiao Fan1, Jin-Hai Li2, Jie-Peng Yao1
1College of Information and Electrical Engineering, China Agricultural University, Beijing, China.
Abstract:
Using PEDOT as the conductive polymer, an innovative small-scale sensor for directly measuring salicylate ions in plants was developed, which avoided the complicated sample pretreatment of traditional analytical methods and realized the rapid detection of salicylic acid. The results demonstrate that this all-solid-state potentiometric salicylic acid sensor is easy to miniaturize, has a longer lifetime (≥1 month), is more robust, and can be directly used for the detection of salicylate ions in real samples without any additional pretreatment. The developed sensor has a good Nernst slope (63.6 ± 0.7 mV/decade), the linear range is 10-2 ~ 10-6 M, and the detection limit can reach (2.8 × 10-7 M). The selectivity, reproducibility, and stability of the sensor were evaluated. The sensor can perform stable, sensitive, and accurate in situ measurement of salicylic acid in plants, and it is an excellent tool for determining salicylic acid ions in plants in vivo.
Related Concept Videos
Tonicity in Plants
Tonicity
Tonicity describes the capacity of a cell to lose or gain water depending on the solute...
Potentiometry: Membrane Electrodes
Potentiometry: Overview
Key Elements for Plant Nutrition
Responses to Salt Stress