Kanil Ranjith Kumar1, Lipika Soni1, Renu Sinha1
1Department of Anaesthesiology, Pain Medicine and Critical Care, All India Institute of Medical Sciences, New Delhi, India.
Background:
Ambu AuraGain has proven to be better compared with other supraglottic airway devices in terms of higher first-attempt insertion success rate, time and ease of insertion, high oropharyngeal leak pressure, and fewer complications in children. The performance of the BlockBuster laryngeal mask has not been evaluated in children.
Aims:
The primary objective of this study was to compare the oropharyngeal leak pressure of the BlockBuster laryngeal mask with those of the Ambu AuraGain during controlled ventilation in children.
Methods:
Fifty children aged 6 months to 12 years with normal airways were randomized into group A (Ambu AuraGain) and group B (BlockBuster laryngeal mask). After administration of general anesthesia, an appropriate size supraglottic airway (size 1.5/2.0/2.5) was inserted according to the groups. Oropharyngeal leak pressure, success and ease of supraglottic airway insertion, gastric tube insertion, and ventilatory parameters were noted. The glottic view was graded by fiberoptic bronchoscopy.
Results:
Demographic parameters were comparable. The mean oropharyngeal leak pressure in the BlockBuster group (24.72 ± 6.81 cm H2 O) was significantly higher than Ambu AuraGain group (17.20 ± 4.28 cm H2 O) by 7.52 cm H2 O (95% CI 4.27 to 10.76; p = 0.001). The mean time for supraglottic airway insertion in the BlockBuster and Ambu AuraGain group was 12.04 ± 2.55 s and 13.64 ± 2.76 s, respectively (mean difference- 1.6 s, 95% CI 0.09-3.12; p = 0.04). Ventilatory parameters, first-attempt supraglottic airway insertion success rate, and ease of gastric tube insertion were comparable between the groups. The BlockBuster group showed easy supraglottic airway insertion compared with the Ambu AuraGain group. The BlockBuster group had better glottic views with only the larynx seen in 23 out of 25 children compared to the Ambu AuraGain with only the larynx seen in 19 out of 25 children. No complication was noted in either group.
Conclusions:
We found that the BlockBuster laryngeal mask has higher oropharyngeal leak pressure compared with Ambu AuraGain in a pediatric population.
More Related Videos
Related Concept Videos
Cardiopulmonary Resuscitation II: ACLS Airway Management
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
General Anesthesia: Overview
General anesthesia induces unconsciousness in the whole body, while the others target specific areas or sensations. It is administered to minimize adverse effects, maintain...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...