Reliability and Validity of A Novel Device for Evaluating the Cervical Proprioception
Meng Zhang1,2, Xiao-Yan Chen3, Sheng-Yu Fu4,1
1Postgraduate School, Dalian Medical University, Dalian, 116000, Liaoning, China.
Pain and Therapy
|March 3, 2023
Summary
The WitMotion sensor shows good reliability and validity for assessing cervical proprioception, offering a practical alternative to traditional methods like laser pointer devices for clinical evaluation.
Area of Science:
- Biomedical Engineering
- Rehabilitation Science
- Clinical Biomechanics
Background:
- Cervical proprioception is crucial for head and neck function.
- Traditional methods for evaluating cervical joint position error (JPE) include laser pointer devices (LPD) and cervical range-of-motion (CROM) instruments.
- There is a need for more advanced, cost-effective, and practical tools for assessing cervical proprioception.
Purpose of the Study:
- To analyze the reliability and validity of the WitMotion sensor (WS) for evaluating cervical proprioception.
- To explore the WS as a cheaper, more convenient, and practical alternative for clinical use.
Main Methods:
- Twenty-eight healthy participants (age 25-66) were recruited.
- Cervical joint position error (JPE) was measured using both the WitMotion sensor (WS) and a laser pointer device (LPD).
- Intra- and inter-rater reliability were assessed using intraclass correlation coefficients (ICC), and validity was analyzed using ICC and Spearman's correlation.
Main Results:
- Intra-rater reliability of the WS was higher than LPD for flexion, right lateral flexion, and left rotation.
- LPD showed higher intra-rater reliability than WS for extension, left lateral flexion, and right rotation.
- Inter-rater reliability for WS and LPD was generally good (ICC > 0.70), with exceptions in extension and left lateral flexion. Validity (ICC > 0.614) was moderate to good for both devices across all movements.
Conclusions:
- The WitMotion sensor demonstrates acceptable to high reliability and validity for assessing cervical joint position error.
- The WS is a viable and practical alternative tool for evaluating cervical proprioception in clinical settings.
- This novel device offers a potentially more accessible option compared to existing methods.