Related Experiment Video
Updated: Aug 8, 2025

Non-fluoroscopic Catheter Tracking for Fluoroscopy Reduction in Interventional Electrophysiology
Published on: May 26, 2015
Ablation of CTI-dependent flutter using different ablation technologies: acute and long-term outcome from the
Giampiero Maglia1, Francesco Pentimalli2, Roberto Verlato3,4
1Cardiology - Coronary Care Unit, Electrophysiology and Cardiac Pacing, Pugliese - Ciaccio Hospital, Via Pio X, 88100, Catanzaro, Italy. pmaglia@libero.it.
Background:
A novel ablation catheter has been released to map and ablate the cavo-tricuspid isthmus (CTI) in patients with atrial flutter (AFL), improving ablation efficiency.
Methods:
We evaluated the acute and long-term outcome of CTI ablation aiming at bidirectional conduction block (BDB) in a prospective, multicenter cohort study enrolling 500 patients indicated for typical AFL ablation. Patients were grouped on the basis of the AFL ablation method (linear anatomical approach, Conv group n = 425, or maximum voltage guided, MVG group, n = 75) and ablation catheter (mini-electrodes technology, MiFi group, n = 254, or a standard 8-mm ablation catheter, BLZ group, n = 246).
Results:
Complete BDB according to both validation criteria (sequential detailed activation mapping or mapping only the ablation site) was achieved in 443 patients (88.6%). The number of RF applications needed to achieve BDB was lower in the MiFi MVG group vs both the MiFi Conv group and the BLZ Conv group (3.2 ± 2 vs 5.2 ± 4 vs 9.3 ± 5, p < 0.0001 for all comparisons). Fluoroscopy time was similar among groups, whereas we observed a reduction in the procedure duration from the BLZ Conv group (61.9 ± 26min) to the MiFi MVG group (50.6 ± 17min, p = 0.048). During a mean follow-up of 548 ± 304 days, 32 (6.2%) patients suffered an AFL recurrence. No differences were found according to BDB achieved by both validation criteria.
Conclusions:
Ablation was highly effective in achieving acute CTI BDB and long-term arrhythmia freedom irrespective of the ablation strategy or the validation criteria for CTI chosen by the operator. The use of an ablation catheter equipped with mini-electrodes technology seems to improve ablation efficiency.
Clinical Trial Registration:
Atrial Flutter Ablation in a Real World Population. (LEONARDO).
Clinicaltrials:
gov Identifier: NCT02591875.