Magnetic Resonance Imaging Quantification of Pulmonary Perfusion using Calibrated Arterial Spin Labeling
Published on: May 30, 2011
Vascular space occupancy asymmetric spin echo (VASO-ASE) for non-invasive quantification of cerebral oxygen
Spencer L Waddle1, Maria Garza1, Chunwei Ying2
1Department of Neurology, Vanderbilt University Medical Center, Nashville, Tennessee, USA.
Purpose:
Asymmetric spin echo (ASE) MRI is a method for measuring regional oxygen extraction fraction (OEF); however, extravascular tissue models have been shown to under-estimate OEF. The hypothesis investigated here is that the addition of a vascular-space-occupancy (VASO) pre-pulse will more fully suppress blood water signal and provide global OEF values more consistent with physiological expectation and 15 O positron emission tomography (PET)-validated T2 -relaxation-under-spin-tagging (TRUST) OEF measures.
Methods:
Healthy adults (n = 14; age = 27.7 ± 5.2 y; sex = 7/7 male/female) were scanned at 3.0T. Multi-echo ASE without inter-readout refocusing (ASERF- ), multi-echo ASE with inter-readout refocusing (ASERF+ ), and single-echo VASO-ASE were acquired twice each with common spatial resolution = 3.44 × 3.44 × 3.0 mm and τ = 0-20 ms (interval = 0.5 ms). TRUST was acquired twice sequentially for independent global OEF assessment (τCPMG = 10 ms; effective TEs = 0, 40, 80, and 160 ms; spatial resolution = 3.4 × 3.4 × 5 mm). OEF intraclass-correlation-coefficients (ICC), summary statistics, and group-wise differences were assessed (Wilcoxon rank-sum; significance: two-sided p < 0.05).
Results:
ASERF+ (OEF = 36.8 ± 1.9%) and VASO-ASE (OEF = 34.4 ± 2.3%) produced OEF values similar to TRUST (OEF = 36.5 ± 4.6%, human calibration model; OEF = 32.7 ± 4.9%, bovine calibration model); however, ASERF- yielded lower OEF (OEF = 26.1 ± 1.0%; p < 0.01) relative to TRUST. VASO-ASE (ICC = 0.61) yielded lower ICC compared to other ASE variants (ICC >0.89).
Conclusion:
VASO-ASE and TRUST provide similar OEF values; however, VASO-ASE spatial coverage and repeatability improvements are required.
More Related Videos
08:35Oxygenation-sensitive Cardiac MRI with Vasoactive Breathing Maneuvers for the Non-invasive Assessment of Coronary Microvascular Dysfunction
Published on: August 17, 2022
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...