False-negative cerebral spinal fluid cryptococcal antigen lateral flow assay due to postzone phenomenon in a patient
Lu Xu1, Yi-Jun Mo2, Xing-Bei Weng2
1Department of Laboratory Medicine, The Affiliated People's Hospital of Ningbo University, Ningbo, Zhejiang Province, China.
Abstract:
This report presents the case of false-negative cerebral spinal fluid (CSF) cryptococcal antigen (CrAg) lateral flow assay (LFA) in a HIV-positive 25-year-old male. The patient presented with headache, nausea and vomiting for 5 days and syncope for 1 day. An initial CSF CrAg LFA test was negative, but a 1:4 dilution of the CSF was weakly positive and a 1:8 dilution was positive. A serum cryptococcal antigen test was weakly positive. Cultures of blood and CSF were all positive for Cryptococcus neoformans. The explanation for the false-negative CSF CrAg LFA test is that the antigen concentration was too high causing the postzone phenomenon.