Photon-Counting Computed Tomography (PC-CT) of the spine: impact on diagnostic confidence and radiation dose
Alexander Rau1,2, Jakob Straehle3, Thomas Stein4
1Department of Diagnostic and Interventional Radiology, University Hospital, Hugstetter Straße 55, 79106, Freiburg, Germany. alexander.rau@uniklinik-freiburg.de.
Objectives:
Computed tomography (CT) is employed to evaluate surgical outcome after spinal interventions. Here, we investigate the potential of multispectral photon-counting computed tomography (PC-CT) on image quality, diagnostic confidence, and radiation dose compared to an energy-integrating CT (EID-CT).
Methods:
In this prospective study, 32 patients underwent PC-CT of the spine. Data was reconstructed in two ways: (1) standard bone kernel with 65-keV (PC-CTstd) and (2) 130-keV monoenergetic images (PC-CT130 keV). Prior EID-CT was available for 17 patients; for the remaining 15, an age-, sex-, and body mass index-matched EID-CT cohort was identified. Image quality (5-point Likert scales on overall, sharpness, artifacts, noise, diagnostic confidence) of PC-CTstd and EID-CT was assessed by four radiologists independently. If metallic implants were present (n = 10), PC-CTstd and PC-CT130 keV images were again assessed by 5-point Likert scales by the same radiologists. Hounsfield units (HU) were measured within metallic artifact and compared between PC-CTstd and PC-CT130 keV. Finally, the radiation dose (CTDIvol) was evaluated.
Results:
Sharpness was rated significantly higher (p = 0.009) and noise significantly lower (p < 0.001) in PC-CTstd vs. EID-CT. In the subset of patients with metallic implants, reading scores for PC-CT130 keV revealed superior ratings vs. PC-CTstd for image quality, artifacts, noise, and diagnostic confidence (all p < 0.001) accompanied by a significant increase of HU values within the artifact (p < 0.001). Radiation dose was significantly lower for PC-CT vs. EID-CT (mean CTDIvol: 8.83 vs. 15.7 mGy; p < 0.001).
Conclusions:
PC-CT of the spine with high-kiloelectronvolt reconstructions provides sharper images, higher diagnostic confidence, and lower radiation dose in patients with metallic implants.
Key Points:
• Compared to energy-integrating CT, photon-counting CT of the spine had significantly higher sharpness and lower image noise while radiation dose was reduced by 45%. • In patients with metallic implants, virtual monochromatic photon-counting images at 130 keV were superior to standard reconstruction at 65 keV in terms of image quality, artifacts, noise, and diagnostic confidence.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Radiological Investigation I: X-ray and CT
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
X-ray Imaging

![Quantitative [18F]-Naf-PET-MRI Analysis for the Evaluation of Dynamic Bone Turnover in a Patient with Facetogenic Low Back Pain](/_next/image?url=https%3A%2F%2Fcloudfront.jove.com%2FCDNSource%2Fteasers%2F58491.jpg&w=3840&q=50)