Evaluation of Computer-Aided Detection During Colonoscopy in the Community (AI-SEE): A Multicenter Randomized
Mike T Wei1, Uday Shankar2, Russell Parvin2
1Stanford University, Stanford, California, USA.
Introduction:
There has been increasing interest in artificial intelligence in gastroenterology. To reduce miss rates during colonoscopy, there has been significant exploration in computer-aided detection (CADe) devices. In this study, we evaluate the use of CADe in colonoscopy in community-based, nonacademic practices.
Methods:
Between September 28, 2020, and September 24, 2021, a randomized controlled trial (AI-SEE) was performed evaluating the impact of CADe on polyp detection in 4 community-based endoscopy centers in the United States Patients were block-randomized to undergoing colonoscopy with or without CADe (EndoVigilant). Primary outcomes measured were adenomas per colonoscopy and adenomas per extraction (the percentage of polyps removed that are adenomas). Secondary end points included serrated polyps per colonoscopy; nonadenomatous, nonserrated polyps per colonoscopy; adenoma and serrated polyp detection rates; and procedural time.
Results:
A total of 769 patients were enrolled (387 with CADe), with similar patient demographics between the 2 groups. There was no significant difference in adenomas per colonoscopy in the CADe and non-CADe groups (0.73 vs 0.67, P = 0.496). Although the use of CADe did not improve identification of serrated polyps per colonoscopy (0.08 vs 0.08, P = 0.965), the use of CADe increased identification of nonadenomatous, nonserrated polyps per colonoscopy (0.90 vs 0.51, P < 0.0001), resulting in detection of fewer adenomas per extraction in the CADe group. The adenoma detection rate (35.9 vs 37.2%, P = 0.774) and serrated polyp detection rate (6.5 vs 6.3%, P = 1.000) were similar in the CADe and non-CADe groups. Mean withdrawal time was longer in the CADe group compared with the non-CADe group (11.7 vs 10.7 minutes, P = 0.003). However, when no polyps were identified, there was similar mean withdrawal time (9.1 vs 8.8 minutes, P = 0.288). There were no adverse events.
Discussion:
The use of CADe did not result in a statistically significant difference in the number of adenomas detected. Additional studies are needed to better understand why some endoscopists derive substantial benefits from CADe and others do not. ClinicalTrials.gov number: NCT04555135.
More Related Videos
15:49Flexible Colonoscopy in Mice to Evaluate the Severity of Colitis and Colorectal Tumors Using a Validated Endoscopic Scoring System
Published on: October 16, 2013
03:44Structured Approach to Colonoscopy Technique Optimization: A Single-Center Experience with Novice Endoscopists
Published on: July 11, 2025
Related Concept Videos
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and...
Endoscopic Procedures II: Colonoscopy
Assessment of the Rectum and Anus
Rectal Inspection
Begin by inspecting the perianal and anal areas for color, texture, rashes,...
Endoscopic Procedures IV: Sigmoidoscopy and Laproscopy
Sigmoidoscopy
Sigmoidoscopy is a diagnostic procedure that uses a flexible sigmoidoscope equipped with a light source and camera to examine the rectum and sigmoid colon. The procedure involves inserting the tube through the anus...
Endoscopic Procedures III: Video Capsule Endoscopy