Effects of ultrasound with an automatic vessel detection system using artificial intelligence on the selection of
Mari Abe-Doi1,2, Ryoko Murayama2,3, Toshiaki Takahashi1,4
1Department of Gerontological Nursing/Wound Care Management, Graduate School of Medicine, The University of Tokyo, Bunkyo-ku, Tokyo, Japan.
Background:
Ultrasound guidance increases the success rate of peripheral intravenous catheter placement. However, the longer time required to obtain ultrasound-guided access poses difficulties for ultrasound beginners. Notably, interpretation of ultrasonographic images is considered as one of the main reasons of difficulty in using ultrasound for catheter placement. Therefore, an automatic vessel detection system (AVDS) using artificial intelligence was developed. This study aimed to investigate the effectiveness of AVDS for ultrasound beginners in selecting puncture points and determine suitable users for this system.
Methods:
In this crossover experiment involving the use of ultrasound with and without AVDS, we enrolled 10 clinical nurses, including 5 with some experience in peripheral intravenous catheterization using ultrasound-aided methods (categorized as ultrasound beginners) and 5 with no experience in ultrasound and less experience in peripheral intravenous catheterization using conventional methods (categorized as inexperienced). These participants chose two puncture points (those with the largest and second largest diameter) as ideal in each forearm of a healthy volunteer. The results of this study were the time required for the selection of puncture points and the vein diameter of the selected points.
Results:
Among ultrasound beginners, the time required for puncture point selection in the right forearm second candidate vein with a small diameter (<3 mm) was significantly shorter when using ultrasound with AVDS than when using it without AVDS (mean, 87 vs 247 s). Among inexperienced nurses, no significant difference in the time required for all puncture point selections was found between the use of ultrasound with and without AVDS. In the vein diameter, significant difference was shown only in the absolute difference at left second candidate among inexperienced participants.
Conclusion:
Ultrasonography beginners needed less time to select the puncture points in a small diameter vein using ultrasound with AVDS than without AVDS.
Related Concept Videos
Ultrasound I: Abdominal Ultrasonography
Abdominal ultrasonography, commonly known as abdominal ultrasound, is a vital, non-invasive medical imaging technique widely used in healthcare.
Procedure:
This diagnostic tool allows the clinician to visually inspect internal structures within the abdomen, including vital organs such as the liver, gallbladder, pancreas, kidneys, and spleen.
The abdominal ultrasound process begins with applying a special gel to the patient's skin over the abdomen. This gel enhances the...
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: