Evaluation of the heart sounds in children using a Doppler Phonolyser
Mohammad Reza Khalilian1, Mahsa Safari2, Mahmoud Hajipour3
1Department of Pediatrics, Shahid Beheshti University of Medical Sciences, Shahid Modarres Educational Hospital, Intersection of Saadat Abad and Yadegar Imam Highway, Tehran, Iran.
Background:
Heart auscultation is an easy and inexpensive tool for early diagnosis of congenital heart defects. In this regard, a simple device which can be used easily by physicians for heart murmur detection will be very useful. The current study was conducted to evaluate the validity of a Doppler-based device named "Doppler Phonolyser" for the diagnosis of structural heart diseases in pediatric patients. In this cross-sectional study, 1272 patients under 16 years who were referred between April 2021 and February 2022, to a pediatric cardiology clinic in Mofid Children Hospital, Tehran, Iran, were enrolled. All the patients were examined by a single experienced pediatric cardiologist using a conventional stethoscope at the first step and a Doppler Phonolyser device at the second step. Afterward, the patient underwent trans-thoracic echocardiography, and the echocardiogram results were compared with the conventional stethoscope as well as the Doppler Phonolyser findings.
Results:
Sensitivity of the Doppler Phonolyser for detecting congenital heart defects was 90.5%. The specificity of the Doppler Phonolyser in detecting heart disease was 68.9% in compared with the specificity of the conventional stethoscope, which was 94.8%. Among the most common congenital heart defects in our study population, the sensitivity of the Doppler Phonolyser was 100% for detection of tetralogy of Fallot (TOF); In contrast, sensitivity of both the conventional stethoscope and the Doppler Phonolyser was relatively low for detecting atrial septal defect.
Conclusions:
Doppler Phonolyser could be useful as a diagnostic tool for the detection of congenital heart defects. The main advantages of the Doppler Phonolyser over the conventional stethoscope are no need for operator experience, the ability to distinguish innocent murmurs from the pathologic ones and no effect of environmental sounds on the performance of the device.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Heart Sounds
Auscultation is the process of listening to these internal body sounds using a stethoscope. The heart produces four types of sounds, but only two—S1 and S2—can usually be heard with a stethoscope.
S1, also known as the "lub" sound, is caused by the closure of atrioventricular (A-V)...
Assessment of the Cardiovascular System IV: Auscultation
Normal Heart Sounds
S1 (First Heart Sound)-
S1 is made by the closure of the mitral and tricuspid valves (atrioventricular valves), marking the beginning of systole.
S2 (Second Heart Sound)-
S2 is made by the closure of the aortic and pulmonic valves (semilunar valves), marking the end of the systole.
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for:
Special considerations while measuring pulse