Impact of Evolution of Self-Expandable Aortic Valve Design: Peri-Operative and Short-Term Outcomes
Evangelia Bei1, Vasileios Voudris2, Konstantinos Kalogeras3
1First Department of Cardiology, Hippokration Hospital, Medical School, National and Kapodistrian University of Athens, 11528 Athens, Greece.
Abstract:
A few data exist on the differences of implantable aortic valve bio-prostheses. We investigate three generations of self-expandable aortic valves in terms of the outcomes. Patients undergoing transcatheter aortic valve implantation (TAVI) were allocated into three groups according to the valve type: group A (CoreValveTM), group B (EvolutTMR) and group C (EvolutTMPRO). The implantation depth, device success, electrocardiographic parameters, need for permanent pacemaker (PPM), and paravalvular leak (PVL) were assessed. In the study, 129 patients were included. The final implantation depth did not differ among the groups (p = 0.07). CoreValveTM presented greater upward jump of the valve at release (2.88 ± 2.33 mm vs. 1.48 ± 1.09 mm and 1.71 ± 1.35 mm, for groups A, B, and C, respectively, p = 0.011). The device success (at least 98% for all groups, p = 1.00) and PVL rates (67% vs. 58%, vs. 60% for groups A, B, and C, respectively, p = 0.64) did not differ. PPM implantation within 24 h (33% vs. 19% vs. 7% for groups A, B, and C, respectively, p = 0.006) and until discharge (group A: 38% vs. group B: 19% and group C: 9%, p = 0.005) was lower in the newer generation valves. Newer generation valves present better device positioning, more predictable deployment, and fewer rates of PPM implantation. No significant difference in PVL was observed.
More Related Videos
14:14Standardized Technique of Aortic Valve Re-implantation for Valve-sparing Aortic Root Replacement
Published on: December 11, 2017
05:31Transcatheter Pulmonary Valve Replacement from Autologous Pericardium with a Self-Expandable Nitinol Stent in an Adult Sheep Model
Published on: June 8, 2022
Related Concept Videos
Aortic Regurgitation I: Introduction
Aortic Regurgitation III: Medical Management
Aneurysm IV: Nursing Management