A Novel Catheter Distal Contact Force Sensing for Cardiac Ablation Based on Fiber Bragg Grating with Temperature
Yuyang Lou1,2, Tianyu Yang2, Dong Luo2
1School of Physics and Electronic Engineering, Chongqing Normal University, Chongqing 401331, China.
Objective:
To accurately achieve distal contact force, a novel temperature-compensated sensor is developed and integrated into an atrial fibrillation (AF) ablation catheter.
Methods:
A dual elastomer-based dual FBGs structure is used to differentiate the strain on the two FBGs to achieve temperature compensation, and the design is optimized and validated by finite element simulation.
Results:
The designed sensor has a sensitivity of 90.5 pm/N, resolution of 0.01 N, and root-mean-square error (RMSE) of 0.02 N and 0.04 N for dynamic force loading and temperature compensation, respectively, and can stably measure distal contact forces with temperature disturbances.
Conclusion:
Due to the advantages, i.e., simple structure, easy assembly, low cost, and good robustness, the proposed sensor is suitable for industrial mass production.