Accuracy of a noninvasive estimated continuous cardiac output measurement under different respiratory conditions: a
Masashi Takakura1,2, Tasuku Fujii3, Tomoya Taniguchi1,2
1Department of Anesthesiology, Nagoya University Hospital, 65 Tsurumai-cho, Showa-ku, Nagoya, 466-8550, Japan.
Purpose:
The estimated continuous cardiac output (esCCO) system was recently developed as a noninvasive hemodynamic monitoring alternative to the thermodilution cardiac output (TDCO). However, the accuracy of continuous cardiac output measurements by the esCCO system compared to TDCO under different respiratory conditions remains unclear. This prospective study aimed to assess the clinical accuracy of the esCCO system by continuously measuring the esCCO and TDCO.
Methods:
Forty patients who had undergone cardiac surgery with a pulmonary artery catheter were enrolled. We compared the esCCO with TDCO from mechanical ventilation to spontaneous respiration through extubation. Patients undergoing cardiac pacing during esCCO measurement, those receiving treatment with an intra-aortic balloon pump, and those with measurement errors or missing data were excluded. In total, 23 patients were included. Agreement between the esCCO and TDCO measurements was evaluated using Bland-Altman analysis with a 20 min moving average of the esCCO.
Results:
The paired esCCO and TDCO measurements (939 points before extubation and 1112 points after extubation) were compared. The respective bias and standard deviation (SD) values were 0.13 L/min and 0.60 L/min before extubation, and - 0.48 L/min and 0.78 L/min after extubation. There was a significant difference in bias before and after extubation (P < 0.001); the SD before and after extubation was not significant (P = 0.315). The percentage errors were 25.1% before extubation and 29.6% after extubation, which is the criterion for acceptance of a new technique.
Conclusion:
The accuracy of the esCCO system is clinically acceptable to that of TDCO under mechanical ventilation and spontaneous respiration.
More Related Videos
07:51Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
04:20Integration of Brain Tissue Saturation Monitoring in Cardiopulmonary Exercise Testing in Patients with Heart Failure
Published on: October 1, 2019
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart...