High-performance liquid chromatography with fluorescence detection for mycophenolic acid determination in saliva
Joanna Sobiak1, Matylda Resztak2, Joanna Banasiak2
1Department of Physical Pharmacy and Pharmacokinetics, Poznan University of Medical Sciences, 3 Rokietnicka Street, 60-806, Poznan, Poland. jsobiak@ump.edu.pl.
Background:
For therapeutic drug monitoring (TDM) of mycophenolic acid (MPA), which is frequently proposed, saliva might be a suitable and easy-to-obtain biological matrix. The study aimed to validate an HPLC method with fluorescence detection for determining mycophenolic acid in saliva (sMPA) in children with nephrotic syndrome.
Methods:
The mobile phase was composed of methanol and tetrabutylammonium bromide with disodium hydrogen phosphate (pH 8.5) at a 48:52 ratio. To prepare the saliva samples, 100 µL of saliva, 50 µL of calibration standards, and 50 µL of levofloxacin (used as an internal standard) were mixed and evaporated to dryness at 45 °C for 2 h. The resulting dry extract was reconstituted in the mobile phase and injected into the HPLC system after centrifugation. Saliva samples from study participants were collected using Salivette® devices.
Results:
The method was linear within the range of 5-2000 ng/mL, was selective with no carry-over effect and met the acceptance criteria for within-run and between-run accuracy and precision. Saliva samples can be stored for up to 2 h at room temperature, for up to 4 h at 4 °C, and for up to 6 months at - 80 °C. MPA was stable in saliva after three freeze-thaw cycles, in dry extract for 20 h at 4 °C, and for 4 h in the autosampler at room temperature. MPA recovery from Salivette® cotton swabs was within the range of 94-105%. The sMPA concentrations in the two children with nephrotic syndrome who were treated with mycophenolate mofetil were within 5-112 ng/mL.
Conclusions:
The sMPA determination method is specific, selective, and meets the validation requirements for analytic methods. It may be used in children with nephrotic syndrome; however further studies are required to investigate focusing on sMPA and the correlation between sMPA and total MPA and its possible contribution to MPA TDM is required.
More Related Videos
08:50Multiplexed Fluorescent Microarray for Human Salivary Protein Analysis Using Polymer Microspheres and Fiber-optic Bundles
Published on: October 10, 2013
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
Related Concept Videos
High-Performance Liquid Chromatography: Types of Detectors
High-Performance Liquid Chromatography: Instrumentation
Gas Chromatography: Types of Detectors-II