Innovated single flush on-line solid-phase extraction in bead injection format for flow programming-based
Petr Chocholouš1, Jan Vinklárek1, Eva Semerádová1
1Department of Analytical Chemistry, Faculty of Pharmacy in Hradec Kralove, Charles University, Hradec Kralove, Czech Republic.
This study introduces an automated on-line solid-phase extraction (SPE) method for precise creatinine determination in urine. The novel flow analysis approach enhances reproducibility and automation for clinical diagnostics.
Area of Science:
- Analytical Chemistry
- Clinical Chemistry
- Biomedical Engineering
Background:
- Automated flow analysis is crucial for reaction-based assays but limited by aggressive reagents damaging manifolds.
- On-line solid-phase extraction (SPE) offers a solution to enhance durability, reproducibility, and automation in flow analysis systems.
Purpose of the Study:
- To develop and validate an automated on-line SPE method for sensitive and selective creatinine determination in human urine using sequential injection analysis.
- To improve the robustness and efficiency of flow analysis for clinical bioanalysis by addressing reagent-induced manifold degradation.
Main Methods:
- Sequential injection analysis coupled with bead injection on-line SPE utilizing a strong cation exchanger.
- UV spectrophotometric detection at 235 nm (with subtraction at 270 nm) for creatinine quantification.
- Automated column packing/disposal, calibration, and a novel zone sequencing technique for accelerated SPE.
Main Results:
- The method achieved high sensitivity and selectivity for creatinine in diluted human urine samples.
- A single run duration was under 3.5 minutes with a relative standard deviation <5.0%.
- Linear calibration range was 1.0-15.0 mmol/L creatinine in urine (R>0.999), with accuracy comparable to routine enzymatic assays.
Conclusions:
- The developed automated on-line SPE method significantly improves flow analysis for creatinine determination, offering enhanced durability and automation.
- The bead injection and zone sequencing techniques accelerate the process and broaden the calibration range while minimizing matrix effects.
- This approach provides a robust, accurate, and efficient alternative for clinical bioanalysis of creatinine in human urine.