Comparison of Three Commercial Methods of Cone-Beam Computed Tomography-Based Dosimetric Analysis of Head-and-Neck
Satyapal Rathee1,2, Benjamin Burke1,2, Amr Heikal1,2
1Department of Oncology, University of Alberta, Edmonton, Alberta, Canada.
Purpose:
This investigation compares three commercial methods of cone-beam computed tomography (CBCT)-based dosimetric analysis to a method based on repeat computed tomography (CT).
Materials And Methods:
Seventeen head-and-neck patients treated in 2020, and with a repeat CT, were included in the analyses. The planning CT was deformed to anatomy in repeat CT to generate a reference plan. Two of the CBCT-based methods generated test plans by deforming the planning CT to CBCT of fraction N using VelocityAI™ and SmartAdapt®. The third method compared directly calculated doses on the CBCT for fraction 1 and fraction N, using PerFraction™. Maximum dose to spinal cord (Cord_dmax) and dose to 95% volume (D95) of planning target volumes (PTVs) were used to assess "need to replan" criteria.
Results:
The VelocityAI™ method provided results that most accurately matched the reference plan in "need to replan" criteria using either Cord_dmax or PTV D95. SmartAdapt® method overestimated the change in Cord_dmax (6.77% vs. 3.85%, P < 0.01) and change in cord volume (9.56% vs. 0.67%, P < 0.01) resulting in increased false positives in "need to replan" criteria, and performed similarly to VelocityAI™ for D95, but yielded more false negatives. PerFraction™ method underestimated Cord_dmax, did not perform any volume deformation, and missed all "need to replan" cases based on cord dose. It also yielded high false negatives using the D95 PTV criteria.
Conclusions:
The VelocityAI™-based method using fraction N CBCT is most similar to the reference plan using repeat CT; the other two methods had significant differences.