Related Experiment Video
Updated: Jul 7, 2026

Tandem High-pressure Freezing and Quick Freeze Substitution of Plant Tissues for Transmission Electron Microscopy
Published on: October 13, 2014
Low-cost stopped-flow and freeze-quench device for double mixing
Łukasz Bujnowicz1, Rafał Pietras1, Marcin Sarewicz1
1Department of Molecular Biophysics, Faculty of Biochemistry, Biophysics and Biotechnology, Jagiellonian University, Kraków, Poland.
Abstract:
Experiments based on fast reagent mixing and observation of reaction progress are considered a powerful tool for investigating the kinetics of chemical and enzymatic reactions. Various spectroscopic methods are used in monitoring the reaction progress, which require different sample preparation methods. Stopped-flow is the most widespread method, where the reaction in the liquid phase is observed by optical absorption spectroscopy. Albeit less popular, the freeze-quench method is also used, in which the reaction is rapidly stopped by freezing the sample at a given time point after the reaction onset. The frozen droplets of the sample are collected and measured at low temperatures in the solid state. Currently, many commercial solutions are available for freeze-quench or stopped-flow experiments, but despite the high price of the devices, most of these do not allow combining both these methods in a single experiment. This study presents a relatively simple solution that combines both these methods, thus making a complete study of chemical or enzymatic reactions possible. Besides, the presented solution enables sequential double mixing of reagents, which is generally problematic and cannot be done using commercial instruments.

