Validation of Blood Pressure Devices as per 2020 World Health Organization Technical Specifications: A Scoping Review
Shruti Murthy1, Eti Rajwar2, Rosemol Johnson2
1The George Institute for Global Health, Hyderabad, India (S.M.).
Background:
To identify and summarize the global research literature on validation of automated noninvasive blood pressure measurement devices (BPMDs) with upper arm cuff, develop a repository of validated BPMDs in compliance with the 2020 World Health Organization technical specifications, and identify challenges and gaps in evidence base on validated BPMDs.
Methods:
A scoping review was conducted. Primary research validating BPMDs complying with the 2020 World Health Organization technical specifications (ie, semiautomated/automated noninvasive devices with upper arm cuff), published in English between January 2000 and December 2021, was included. We searched MEDLINE, Web of Science, Scopus, EMBASE, CINAHL, CENTRAL, ProQuest and the dabl website.
Results:
We included 269 studies validating 251 BPMDs across 89 manufacturers. Omron (29%), Microlife (10%), and A&D Company (8%) were the top 3 manufacturers. The 3 most frequently used validation protocols were the European Society of Hypertension-international protocol 2002 (27%), European Society of Hypertension-international protocol 2010 (25%), and modified British Hypertension Society protocol 1993 (16%), respectively. Nearly 45% of the validated BPMDs were intended for use in clinical settings, 38% were for home or self-measurement use, and 48% were for general adults. Most studies reported that BPMDs passed the validation criteria. There was inadequate reporting across studies, especially pertaining to validation settings.
Conclusions:
Most BPMDs fulfilled the validation criteria. However, there are considerable gaps in BPMD research in terms of geographical representation, including specific target populations and diseases/conditions, and a range of arm circumferences. Additionally, a potential strategy is required to accelerate the adoption of the Association for the Advancement of Medical Instrumentation (AAMI)/European Society of Hypertension/International Organization for Standardization Universal Standard (International Organization for Standardization 81060-2:2018) for BPMD validation.
More Related Videos
14:09Tilt Testing with Combined Lower Body Negative Pressure: a "Gold Standard" for Measuring Orthostatic Tolerance
Published on: March 21, 2013
05:51Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Pre-Procedural Guidelines for Assessing Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure