Virtual monochromatic imaging with projection-based material decomposition algorithm for metal artifacts reduction in
Chia-Hao Chang1,2, Hsiang-Ning Wu1, Ching-Han Hsu2
1Health Physics Division, Institute of Nuclear Energy Research, Atomic Energy Council, Taoyuan, Taiwan.
Abstract:
Metal artifacts present a major challenge to computed tomography (CT) because they reduce the image quality in medical diagnosis and treatment. Several metal artifact reduction (MAR) methods have been proposed to address this issue in previous studies. This study aimed to synthesize a virtual monochromatic image for MAR in CT images using projection-based material decomposition (MD) algorithms. We developed a spectral micro-CT prototype system equipped with a photon-counting detector (PCD) and PCD-CT imaging simulator to assess the performances of different MAR methods. Two projection-based MD algorithms were implemented and evaluated for their MAR performances in CT images and compared with conventional sinogram inpainting MAR methods. Different parts of digital 4D-extended cardiac torso (XCAT) phantoms with metal implants were designed to simulate various real scenarios. A homemade metal artifact evaluation (MAE) phantom was used to evaluate the MAR performance in experiments. The simulated results of the XCAT phantom indicated that the projection-based virtual monochromatic CT (VMCT) images provided better image quality than the conventional MAR images without blurring the normal tissues at the position of the metal artifacts. Various quantitative indicators support this conclusion. Additionally, the experimental results of the MAE phantom reveal that projection-based VMCT images can avoid image distortion caused by metal artifacts, unlike conventional MAR methods. In regards to the projection-based VMCT images, the simulated and experimental results demonstrated that using the linear maximum likelihood estimators with an error correction look-up table algorithm yielded better MAR performance compared to that obtained using a polynomial algorithm. Furthermore, projection-based VMCT images can not only reduce metal artifacts effectively but also simultaneously prevents object blurring at the metal artifact position and image distortion of the metal implants. Hence, the CT image quality can be further improved to increase the abilities for both preoperative and postoperative assessment of metal implants.
More Related Videos
09:47Array Tomography Workflow for the Targeted Acquisition of Volume Information using Scanning Electron Microscopy
Published on: July 15, 2021
10:10Three-Dimensional Particle Shape Analysis Using X-ray Computed Tomography: Experimental Procedure and Analysis Algorithms for Metal Powders
Published on: December 4, 2020
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
Imaging Studies III: Computed Tomography
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...