Related Experiment Video
Updated: Sep 25, 2026

Enabling High Grayscale Resolution Displays and Accurate Response Time Measurements on Conventional Computers
Published on: February 29, 2012
Color performance of video endoscopes: quantitative measurement of color reproduction
K Knyrim1, H K Seidlitz, F Hagenmüller
1II. Med. Klinik, Klinikum rechts der Isar, TUM, Neuherberg, West-Germany.
Abstract:
We have conducted a study aimed at testing the color performance of video endoscopes in quantitative terms for the first time. The video endoscopes investigated were from Fuji, Olympus, Toshiba/Machida, and Welch-Allyn. The tests were carried out with an opto-electronic color analyser using standardized color charts and a neutral grey chart. Hue and chroma of the reproduced colors were measured. The Olympus video endoscope reproduces the hue very well, showing only slight color desaturation. The Fuji video endoscope had a yellowish tint, whereas the picture obtained with the WA video endoscope produced an inhomogeneous picture with a pronounced red/purplish cast at the top of the screen and almost neutral reproduction at the lower half. The Toshiba video endoscope was the only one with a color chip, and reproduced hue fairly well, but its colors were weakly saturated.

