Related Experiment Video
Updated: Aug 6, 2025

Using an Ingestible Telemetric Temperature Pill to Assess Gastrointestinal Temperature During Exercise
Published on: October 7, 2015
Heat stroke burden and validity of wearable-derived core temperature estimation during elite military training
Courtney Hintz1, Danielle M Presley2, Cody R Butler1
1Special Warfare Human Performance Support Group, USAF, Lackland AFB, San Antonio, TX, USA.
Objectives:
Exertional heat stroke (EHS) remains a significant health concern while training in hot environments, prompting the development of noninvasive wearable technology for monitoring estimated core temperature (ECT). The objective of this study was to review the effectiveness of an ECT device during elite military training by providing a case series of individuals who developed EHS.
Methods:
This is a retrospective study of EHS cases during high-intensity training while wearing the ECT device, Zephyr BioharnessTM. Data was collected from January 2021 through September 2021 at the Air Force Special Warfare Training pipeline in San Antonio, TX. Rectal temperatures of EHS diagnoses, defined by central nervous system (CNS) dysfunction and rectal temperature approaching or >40°C (104°F), were compared to ECT reading via Zephyr BioharnessTM. Incidence rates and psychometric properties were calculated using R package.
Results:
A total of 47,058 daily peak ECT measurements were collected among 1,364 trainees. A total of 499 trainees flagged as potential EHS by Zephyr BioharnessTM reading >39.7°C (103.5°F). The incidence of confirmed EHS was 0.8/1000 person-months. Of the 10 confirmed EHS cases (9 males, 1 female; age = 23.4 ± 2.7 yrs; BMI = 25 ± 2; body fat = 13 ± 5%), 8 trainees had a peak ECT reading below 39.7°C which resulted in a sensitivity of 20%, specificity of 98.9%, positive likelihood ratio (LR) of 18.93 (95% Confidence Interval [CI] 5.5-65.6), and a negative LR of 0.81 (95% CI 0.6-1.1).
Conclusion:
ECT had substantial false positive and negative rates. Further studies are needed to validate this technology in other populations, and the algorithm used in this device needs to be refined to better capture the environmental and physical requirements in the special operations population.
Related Concept Videos
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...