Real-world use of Control-IQ™ technology automated insulin delivery in pregnancy: A case series with qualitative
Xinye Serena Wang1, Amy D Dunlop2, Julie A McKeen2
1Department of Obstetrics and Gynaecology, University of Calgary, Calgary, Alberta, Canada.
Background:
Most commercially available automated insulin delivery (AID) systems are not approved for pregnancy use. Information regarding use of the Tandem t:slim X2 insulin pump with Control-IQ™ technology in pregnancy is lacking.
Aims:
This case series aimed to explore glycaemic and qualitative experiences of four early adopters of Control-IQ technology in pregnancy.
Methods:
Participants used Control-IQ technology in pregnancy and postpartum and consented to analysis of glycaemic data and semi-structured interviews.
Results:
Case 1 began Control-IQ technology at 10 weeks gestation. Her pregnancy glucose time-in-range (3.5-7.8 mmol/L [63-140 mg/dL]) increased from 58.7% to 73.3% by third trimester. Cases 2-4 began using Control-IQ technology 0-2 months preconception. Pregnancy time-in-range glucose increased from 73.4% to 78.7%, 78% to 83.6%, and 46.5% to 71.9% between first and third trimesters, respectively. A mid-pregnancy decline in time-in-range glucose was observed in two of the four participants related to suboptimal pump setting adjustments and delays in sensor and infusion set replacement. No diabetic ketoacidosis or severe hypoglycaemia occurred. All participants reported reduced diabetes management burden and improved sleep with Control-IQ technology use.
Conclusions:
Early adopters of Control-IQ technology safely used this system off-label in pregnancy and reported reduced diabetes management burden and improved sleep. The largest glycaemic improvements were observed among those with the lowest pregnancy time-in-range glucose at the beginning of pregnancy. Participants with low pregnancy glucose time-in-range increased their time-in-range with Control-IQ technology use and participants with high pregnancy glucose time-in-range maintained and increased their time-in-range with less diabetes management burden.
Related Concept Videos
Insulin Formulations: Types and Delivery
Short-acting insulins are divided into...
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...
Diabetes Mellitus: Type 2 and Gestational
Insulin: Biosynthesis, Chemistry, and Preparation
Damage or functional impairment of β-cells inhibits insulin production, leading to diabetes. Diabetes treatment...
Hormones Regulating Blood Glucose
In addition to accelerating glucose uptake and utilization, insulin has...
Diabetes: Management and Pharmacotherapy
Insulin remains the cornerstone of treatment for most patients with type 1 and many...