Fabrication of a Dipole-assisted Solid Phase Extraction Microchip for Trace Metal Analysis in Water Samples
Published on: August 7, 2016
Direct magnetic sorbent sampling flame atomic absorption spectrometry (DMSS-FAAS) for highly sensitive determination
Lucimara Mendonça Costa1, Fabrício Alves Borges1, Marcello Henrique da Silva Cavalcanti2
1Laboratory of Toxicant and Drug Analyses, Faculty of Pharmaceutical Sciences, Federal University of Alfenas, 37130-001, Alfenas, MG, Brazil; Institute of Chemistry, Federal University of Alfenas, 37130-001, Alfenas, MG, Brazil.
Abstract:
A procedure of direct magnetic sorbent sampling in flame atomic absorption spectrometry (DMSS-FAAS) was developed in this work. Metal-loaded magnetic sorbents were directly inserted in the flame of the FAAS for direct metal desorption/atomization. Magnetic graphene oxide aerogel (M-GOA) particles were synthesized, characterized, and used as a proof-of-concept in the magnetic dispersive solid phase extraction of Pb2+ ions from water samples. M-GOA was selected because is a light and porous sorbent, with high adsorption capacity, that is quickly burned by the flame. Magnetic particles were directly inserted in the flame by using a metallic magnetic probe, thereby avoiding the need for a chemical elution step. As all the extracted Pb2+ ions arrive to the flame without passing through the nebulization system, a drastic increase in the analytical signal was achieved. The improvement in the sensitivity of the proposed method (DMSS-FAAS) for Pb2+ determination was at least 40 times higher than the conventional procedure in which the Pb2+ is extracted, eluted, and analyzed by conventional flame atomic absorption spectrometry (FAAS) via the nebulization system. The analytical curve was linear from 5.0 to 180.0 μg L-1 and the limit of detection was found to be 1.30 μg L-1. Background measurements were insignificant, and the atomic absorption peaks were narrow and reproducible. Precision assessed as a percentage of the relative standard deviation %RSD was found to be 17.4, 7.1, and 7.8% for 10, 70, and 180 μg L-1 levels, respectively. The method showed satisfactory results even in the presence of other ions (Al3+, Cr3+, Co2+, Cu2+, Fe3+, Mn2+, Ba2+, Mg2+, and Li+). The performance of the new system was also evaluated for Cd2+ ions, as well as by using other magnetic particles available in our lab: magnetic carbon nanotubes (M-CNTs), magnetic restricted access carbon nanotubes (M-RACNT), magnetic poly (methacrylic acid-co-ethylene glycol dimethacrylate) (M-PMA), magnetic nanoparticles coated with orange powder peel (M-OPP), and magnetic nanoparticles covered with SiO2 (M - SiO2). Analytical signals increased for both analytes in all sorbents (increases of about 4-37 times), attesting the high potential and applicability of the proposed method. Simplicity, high analytical frequency, high detectability and reproducibility, low cost, and possibility of being totally mechanized are the most relevant advantages.
Related Concept Videos
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
Atomic Emission Spectroscopy: Overview
Atomic Fluorescence Spectroscopy
Atomic Absorption Spectroscopy: Overview
When irradiated by EMR of a particular wavelength, these...
Atomic Emission Spectroscopy: Lab
High-Performance Liquid Chromatography: Types of Detectors