Claudio de Zuloaga1, Andrés Ferrari2
1Hospital Nacional Profesor Alejandro Posadas, Ecuador 1449 10ª "B" (CABA), Buenos Aires, Argentina.
Background:
The adverse effects of conventional right ventricular (RV) apical pacing prompted the search for more physiological pacing sites, such as selective and nonselective His bundle pacing (HBP), a variant of nonselective HBP (para-Hisian pacing), and mid-septal pacing. However, knowledge of their true benefit on the physiology of ventricular activation, lead stability, and pacing thresholds is limited.
Methods And Results:
We included 152 consecutive patients (mean age 61 ± 24, 63% men) in this retrospective study. Of these, 137 patients with different bradyarrhythmias underwent active fixation lead implantation at the RV apex (n = 54), para-Hisian region (n = 66), or mid interventricular septum (n = 17). Fifteen patients with ventricular preexcitation due to an accessory pathway not undergoing pacing were included as controls. A 12‑lead ECG was recorded in all patients, and cardiac electrical synchrony was assessed using the Synchromax® cross-correlation cardiac synchrony index (CSI).
Results:
QRS duration was prolonged in all pacing sites: from 114 ± 28 to 160 ± 29 (RV apex), from 110 ± 28 to 122 ± 29 (para-Hisian), and from 121 ± 24 to 154 ± 30 (mid interventricular septum). The CSI was significantly improved only in patients undergoing para-Hisian pacing, despite a slight widening of the QRS interval. There was no difference in pacing thresholds and sensed R-wave voltage between pacing sites. Only 1 lead, implanted at the para-Hisian region (1.5%), was dislodged towards the mid septum 48 h after implantation but did not require repositioning.
Conclusions:
QRS duration was not associated with changes in CSI, meaning that QRS width does not significantly affect electrical synchrony.