Case Series of Early Structural Valve Deterioration of Trifecta Bioprosthesis - New Zealand Experience
Trevor D Tnay1,2, Lily Kang2, Andrew Mekhail1
1Department of Cardiothoracic Surgery, Wellington Regional Hospital, Capital and Coast District Health Board, Wellington, New Zealand.
Purpose:
Structural valve deterioration (SVD) remains a limitation on the use of bioprosthetic valves, with patient and valve-related factors contributing to early SVD. The Trifecta valve has been reported to have excellent hemodynamics but studies have highlighted early failure. We present a review and case series at a New Zealand tertiary hospital defining early SVD as failure within 3 years of implant.
Methods:
A retrospective review from January 2015 to July 2019 included 525 patients undergoing surgical aortic valve replacement with 263 patients receiving an Abbott Trifecta or Trifecta Glide Technology (GT) valve. Our review found an acceptable safety profile for the valve with excellent hemodynamics, with a low mortality, stroke, and permanent pacemaker rate.
Results:
Three patients out of 263 were identified from the study period as having early SVD requiring reintervention within 3 years of valve implantation leading to a 1.14% failure rate. One of the valves that had early SVD was a new generation Trifecta GT. An additional four patients were identified to have valves implanted prior to the study period and had valve failure at greater than 3 years post implantation. Five cases had cusp tears as their mechanism of failure, raising concerns about durability.
Conclusion:
The Trifecta valve has an acceptable safety profile and offers good hemodynamics due to the externally mounted leaflets. However, our experience of early SVD and failure is concerning for valve durability. Further comparison to other bioprosthetic valves and longer term follow-up are required to characterize the mechanism of failures.
More Related Videos
14:14Standardized Technique of Aortic Valve Re-implantation for Valve-sparing Aortic Root Replacement
Published on: December 11, 2017
05:31Transcatheter Pulmonary Valve Replacement from Autologous Pericardium with a Self-Expandable Nitinol Stent in an Adult Sheep Model
Published on: June 8, 2022
Related Concept Videos
Mitral Valve Prolapse I: Introduction
Mitral Stenosis III: Medical Management