A comparison of the warming capabilities of two Baragwanath rewarming appliances with the Hotline fluid warming
K Wilson1, M Fourtounas1, C Anamourlis1
1Department of Anaesthesiology, Faculty of Health Sciences, University of the Witwatersrand, Johannesburg, South Africa.
Background:
Accidental intraoperative hypothermia is a common and avoidable adverse event of the perioperative period and is associated with detrimental effects on multiple organ systems and postoperative patient outcomes. In a resource-limited environment, prevention of intraoperative hypothermia is often challenging. Resourceful clinicians overcome these challenges through creative devices and frugal innovations.
Objectives:
To investigate the thermal performance of two Baragwanath Rewarming Appliances (BaRA) against that of the Hotline device to describe an optimal setup for these devices.
Methods:
This was a quasi-experimental laboratory study that measured the thermal performance of two BaRA devices and the Hotline device under a number of scenarios. Independent variables including fluid type, flow rate, warming temperature and warming transit distance were sequentially altered and temperatures measured along the fluid stream. Change in temperature (ΔT) was calculated as the difference between entry and exit temperature for each combination of variables for each warming device.
Results:
A total of 219 experiments were performed. At a temperature of 43.0°C and a transit distance of 200 cm, the BaRA A configuration either matched or exceeded the ΔT of the Hotline over all fluid type and flowrate combinations. The BaRA B configuration does not provide comparable thermal performance to the Hotline. Measured flowrates were noticeably slower than manufacturer-quoted values for all intravenous (IV) cannulae used.
Conclusion:
A warm-water bath at 43.0°C with 200 cm of submerged IV tubing provides thermal performance comparable to the Hotline device, with all fluid type and flowrate combinations.
Contributions Of The Study:
The present study provides an evidence-based method for warming intravenous fluid in resource-limited scenarios.
Related Concept Videos
Decreased Body Temperature
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Methods of reducing fever
Pharmacological Methods of Reducing Fever:
Mechanism of heat transfer
Refrigerators and Heat Pumps
A household refrigerator removes heat from...
Heating and Cooling Curves
For instance, the addition of heat raises the temperature of a solid; the amount of heat absorbed depends on the heat capacity of the solid (q = mcsolidΔT). According to thermochemistry, the relation between the amount of heat absorbed or released by a substance, q, and its...