Are portable ankle brachial pressure index measurement devices suitable for hypertension screening?
Justyna Janus1,2, Jennifer K Nicholls1,2, Edward Pallett1,2
1Department of Cardiovascular Sciences, University of Leicester, Leicester, United Kingdom.
Objective:
In a large-scale population cardiovascular screening programme, peripheral artery disease (PAD) and hypertension would ideally be rapidly assessed using a single device. The ankle-brachial pressure index (ABPI) is calculated by comparing the ankle and brachial blood pressure (BP). However, it is currently unclear whether brachial BP measurements provided by automated PAD screening systems are sufficiently accurate for simultaneous hypertension screening.
Methods:
Two portable PAD screening devices, the MESI ABPI MD and Huntleigh's Dopplex ABIlity, were evaluated following the European Society of Hypertension International Protocol (ESH-IP) Revision 2010 using a mercury-free sphygmomanometer as a reference device.
Results:
On average, the MESI slightly underestimated brachial systolic blood pressure (BP) with a bias and standard deviation (SD) of -3.5 (SD: 3.3) mmHg and diastolic BP with a bias of -1.5 (SD: 2.3) mmHg. For systolic BP estimates, the Dopplex was more accurate than the MESI with a lower bias of -0.5 (SD: 4.2) mmHg but less precise. The MESI successfully fulfilled all the requirements of the ESH-IP for hypertension screening. The Dopplex device failed the ESH-IP due to the absence of DBP measurements.
Conclusions:
The MESI device appears to be suitable for simultaneous PAD and hypertension screening as part of a preventative care programme. Huntleigh's Dopplex ABIlity failed to pass the ESH-IP validation test. Further clinical trials are underway to assess the use of the MESI for simultaneous screening for hypertension and PAD in a population screening setting.
Related Concept Videos
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Pre-Procedural Guidelines for Assessing Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.