Related Experiment Video
Updated: Aug 6, 2025

Individualized Stem-positioning in Calcar-guided Short-stem Total Hip Arthroplasty
Published on: February 27, 2018
Micromotion and stress shielding between taper fluted and cylindrical femoral stems for Paprosky bone deficiency type
S Limmahakhun1, N Adulkasem2, C Rungsiyakull3
1Department of Orthopaedic Surgery, Faculty of Medicine, Chiang Mai University, 110 Inthawarorot Rd., Suthep, Muang District, Chiang Mai, 50200, Thailand. sakka_lim@hotmail.com.
Purpose:
Taper fluted and cylindrical femoral stems have been commonly used for revision surgery with femoral bone loss. Shape and material differences between taper fluted and cylindrical confused surgeons whether superiority and inferiority in comparisons especially for type IIIB Paprosky bone deficiency. This study aims to investigate the micromotion between implant and bone interface and stress distribution around the implant in human cadaveric experiments.
Methods:
Paired comparisons of 7 cadavers between taper fluted stem and cylindrical stem were tested under dynamic cyclic loading.
Results:
The results showed appropriate bone-implant micromotions for bone osteointegration could be observed between both taper fluted and cylindrical stems (85 ± 10 μm and 113 ± 80 μm respectively, p = 0.59). Taper fluted stem attained the load distribution from the proximal to distal surrounding bone (from 4.92 ± 2.87 MPa to 2.14 ± 1.43 MPa, p = 0.43), while the cylindrical stem type showed the proximal bone stress shielding (from 2.56 ± 0.76 MPa to 5.23 ± 0.77 MPa, p = 0.03).
Conclusion:
Both taper fluted and cylindrical femoral stems provided an adequate initial stability with a proper micromotions for bone-implant osteointegration for type IIIB Paprosky femoral bone loss. The taper fluted femoral stem had better biomechanical advantage in terms of greater stress transfer to periprosthetic bone.
Related Concept Videos
Bones of the Lower Limb: Femur and Patella
Bones of the Lower Limb: Tibia and Fibula