Performance, safety and efficiency comparison between 10,000 and 5000 cuts per minute vitrectomy using a 25G cutter:
Nicholas S K Fung1,2, Anthony K H Mak1, Marten Brelen3
1Department of Ophthalmology, Grantham Hospital, Hong Kong, China.
International Journal of Retina and Vitreous
|March 22, 2023
Summary
This study found no significant difference in vitrectomy times between 10,000 cuts per minute (cpm) and 5000 cpm vitrectomy probes. Equipment limitations may not be the primary factor affecting surgical efficiency.
Area of Science:
- Ophthalmology
- Surgical Technology
- Medical Devices
Background:
- Vitrectomy surgery is a common ophthalmic procedure.
- The efficiency of vitrectomy probes can impact surgical outcomes.
- Different vitrectomy probe speeds may influence surgical performance.
Purpose of the Study:
- To compare the performance of the 25+® UltraVit® 5000 cuts per minute (cpm) vitrectomy probe against the 25+® UltraVit® 10,000 cpm beveled tip, dual drive vitrectomy probe.
- To evaluate differences in vitrectomy time, posterior vitreous detachment (PVD) induction, intraoperative complications, and instrument usage between two vitrectomy probe speeds.
Main Methods:
- A prospective randomized controlled clinical trial involving 52 eyes of 52 patients.
- Patients were randomized into either a 10,000 cpm (25 patients) or 5000 cpm (27 patients) vitrectomy group.
- Evaluations included preoperative, intraoperative, and postoperative assessments at multiple time points up to 3 months.
Main Results:
- Vitrectomy time was slightly shorter in the 10,000 cpm group (413.7 s) compared to the 5000 cpm group (463.4 s), but this difference was not statistically significant (p=0.5999).
- One iatrogenic retinal break occurred in the 10,000 cpm group, and two occurred in the 5000 cpm group.
- No significant differences were observed in the time for posterior vitreous detachment (PVD) induction or the number of instruments used between the groups.
Conclusions:
- The tested vitrectomy probe speeds (5000 cpm vs. 10,000 cpm) did not yield statistically significant differences in vitrectomy times.
- Surgical efficiency in vitrectomy may be influenced by factors beyond probe speed, such as surgeon technique or other equipment variables.
- Further research may be needed to identify the key determinants of optimal vitrectomy performance.