Related Experiment Video
Updated: Aug 6, 2025

A Filter-based Surface Enhanced Raman Spectroscopic Assay for Rapid Detection of Chemical Contaminants
Published on: February 19, 2016
Rapid on-site detection of persistent organic pollutants using multiwalled carbon nanotube-modified paper spray
Xiangyu Guo1, Junfang Li1, Hua Bai1
1Key Laboratory of Consumer Product Quality Safety Inspection and Risk Assessment for State Market Regulation, Chinese Academy of Inspection and Quarantine, Beijing Economic-Technological Development Area, Beijing, China.
Rationale:
Rapid on-site detection of persistent organic pollutants (POP) is highly desirable for environmental protection.
Methods:
Herein, a rapid on-site analytical workflow was developed for the investigation of polycyclic aromatic hydrocarbons and perfluorinated compounds using multiwalled carbon nanotube-modified paper spray ionization (PSI) coupled with a miniature ion trap mass spectrometer. Critical parameters regarding PSI and miniature mass spectrometry analysis were optimized.
Results:
The analytical performance of the developed method was evaluated under optimized conditions, obtaining a short analysis duration of less than 1 min, sufficient linearity with correlation coefficients greater than 0.99, acceptable recovery rates of 93.1%-105.8% with relative standard deviations of between 3.5% and 10.3%, and reasonable sensitivity with limits of detection and quantitation of 2-200 and 5-500 μg/L, respectively.
Conclusions:
Considering these aspects, it was concluded that the present approach demonstrated a promising solution for rapid on-site detection of emerging POPs.
More Related Videos
07:32Detection of 3-Nitrotyrosine in Atmospheric Environments via a High-performance Liquid Chromatography-electrochemical Detector System
Published on: January 30, 2019
06:43Leaf Spray Mass Spectrometry: A Rapid Ambient Ionization Technique to Directly Assess Metabolites from Plant Tissues
Published on: June 21, 2018
Related Concept Videos
Gas Chromatography: Types of Detectors-II
High-Performance Liquid Chromatography: Types of Detectors