Ventricular fibrillation episodes related to managed ventricular pacing algorithm in a patient with mitral valve
Janet X C Chen1, Jenny Fong1, Han S Lim2
1Austin Health, Victoria, Australia.
Abstract:
We report a case of an implant cardiac defibrillator (ICD) patient who an ICD implanted for ventricular fibrillation (VF) related to mitral valve prolapse. He has 2 episodes of VF in his device lives. First episode of VF in year 2016 was initiated after a pause related to the MVP™ algorithm with a critically timed PVC. MVP™ was turned off which prevented further VF episodes. However, MVP™ was turned back on at the device replacement in 2018. A second VF episode developed with similar mechanism in 2021 and MVP™ was subsequently turned off with no further VF recorded. This case highlights the importance of recognizing the mechanism of initiation of tachy-arrhythmia episodes and serves as an important reminder regarding optimization of device settings at the time of replacement.
More Related Videos
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
07:56A Model of Long-Term Ventricular Fibrillation in Isolated Rat Hearts
Published on: February 17, 2023
Related Concept Videos
Mitral Valve Prolapse II: Assessment and Management
Mitral Valve Prolapse I: Introduction
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Mitral Valve Prolapse III: Nursing Management
Mitral Stenosis III: Medical Management
Mitral Regurgitation III: Medical Management