Frontloading SITA-Faster Can Increase Frequency and Reliability of Visual Field Testing at Minimal Time Cost
Jeremy C K Tan1, Michael Kalloniatis2, Jack Phu3
1Faculty of Medicine and Health, University of New South Wales, Kensington, NSW, Australia; Department of Ophthalmology, Prince of Wales Hospital, Randwick, NSW, Australia.
Purpose:
To report the outcomes of frontloaded visual field (VF) testing (2 tests per eye on the same visit) over 2 longitudinal, consecutive visits using SITA-Faster (SFR) in terms of global indices, reliability metrics, and test duration.
Design:
Prospective longitudinal study.
Subjects:
A total of 902 eyes of 463 subjects with normal, suspect, or manifest glaucoma.
Methods:
Two intravisit SFR VF tests (T1 and T2) per eye at an initial (Ti) and follow-up (Tf) visit.
Main Outcome Measures:
Intra- and intervisit global indices, reliability metrics, and test durations.
Results:
The mean age of the subjects was 63.6 years, and 58.3% were male. Seven hundred ninety eyes (87.4%) had a diagnosis of glaucoma or glaucoma suspicion. The mean duration between visits was 265.0 (standard deviation 98.8) days. In total, 3608 VF tests were analyzed, with the correlation of mean deviation (MD) values of the frontloaded tests at each visit high (T1/T2 MD correlation at initial visit r = 0.83, root mean squared error [RMSE] = 1.26, follow-up visit r = 0.83, RMSE = 1.25, P < 0.0001) and greater than the correlation of MD between visits (Ti1/Tf1 MD correlation r = 0.72, RMSE = 1.31). There was a significant intra-visit decrease in rates of abnormally high sensitivity in the glaucoma hemifield test (3.2% vs. 1.6%, P = 0.0023) and rates of unreliable test results (15.4% vs. 9.2%, P = 0.002) from T1 to T2 in both visits, with a corresponding significant decrease in MD (-1.28 dB vs. -1.68 dB, P < 0.0001) and VF index (P = 0.03). The mean duration of each SFR test was 132.6 (SD 27.2) seconds.
Conclusions:
Frontloading VFs using SFR produced sets of repeatable perimetric data with significant improvement of reliability indices from the first to second test. This may help increase testing frequency at minimal time cost to meet recommended guidelines and for evaluating patients prone to high variability.
Financial Disclosure(S):
The author(s) have no proprietary or commercial interest in any materials discussed in this article.
More Related Videos
06:25Author Spotlight: Assessment of Visual Acuity in Central Vision Loss Through Motion-Based Peripheral Vision Testing
Published on: February 23, 2024
12:23Dynamic Visual Tests to Identify and Quantify Visual Damage and Repair Following Demyelination in Optic Neuritis Patients
Published on: April 14, 2014