Comparison of Blood Pressure Measurements by Currently Available Multiparameter Monitors and Mercury Column
Adil Ahmed Khan1, Pramod Kumar Gupta2, Arun Kumar Baranwal3
1Department of Neonatology, All India Institute of Medical Sciences Jodhpur, Jodhpur, Rajasthan, India.
Background:
The multiparameter monitor (MPM) is replacing mercury column sphygmomanometers (MCS) in acute care settings. However, data on the former's accuracy in critically ill children are scarce and mostly extrapolated from adults. We compared non-invasive blood pressure (NIBP) measurements by MPMs with MCS in pediatric intensive care unit (PICU).
Patients:
Adequately sedated and hemodynamically stabilized children (age, 1-144 months) were prospectively enrolled.
Materials And Methods:
Three NIBP measurements were obtained from MCS (Diamond®, India) and MPM (Intellivue MX800® or Ultraview SL®) in rapid succession in the upper limb resting in supine position. Respective three measurements were averaged to obtain a paired set of NIBP readings, one each from MCS and MPM. Such readings were obtained thrice a day. NIBP readings were then compared, and agreement was assessed.
Results:
From 39 children [median age (IQR), 30 (10-72) months], 1,690 sets of NIBP readings were obtained. A-third of readings were from infants and children >96 months, while 383 (22.6%) readings were from patients on inotropes. Multiparameter monitors gave significantly higher NIBP readings compared to MCS [median systolic blood pressure (SBP), 6.5 (6.4-6.7 mm Hg); diastolic blood pressure (DBP), 4.5 (4.3-4.6 mm Hg); mean arterial pressure (MAP), 5.3 (5.1-5.4 mm Hg); p < 0.05]. It was consistent across age, gender, and critical care characteristics. Multiparameter monitors overestimated SBP in 80% of readings beyond the maximal clinically acceptable difference (MCAD).
Conclusions:
Non-invasive blood pressure readings from MCS and MPMs are not interchangeable; SBP was 6-7 mm Hg higher with the latter. Overestimation beyond MCAD was overwhelming. Caution is required while classifying systolic hypotension with MPMs. Confirmation with auscultatory methods is advisable. More studies are required to evaluate currently available MPMs in different pediatric age groups.
How To Cite This Article:
Khan AA, Gupta PK, Baranwal AK, Jayashree M, Sahoo T. Comparison of Blood Pressure Measurements by Currently Available Multiparameter Monitors and Mercury Column Sphygmomanometer in Patients Admitted in Pediatric Intensive Care Unit. Indian J Crit Care Med 2023;27(3):212-221.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: