Clinical Validation of a Wearable Piezoelectric Blood-Pressure Sensor for Continuous Health Monitoring
Seongwook Min1, Dong Hyun Kim1, Daniel J Joe2
1Department of Materials Science and Engineering, Korea Advanced Institute of Science and Technology (KAIST), 291 Daehak-ro, Yuseong-gu, Daejeon, 34141, Republic of Korea.
Abstract:
Wearable blood-pressure sensors have recently attracted attention as healthcare devices for continuous non-invasive arterial pressure (CNAP) monitoring. However, the accuracy of wearable blood-pressure (BP) monitoring devices has been controversial due to the low signal quality of sensors, the absence of an accurate transfer function to convert the sensor signals into BP values, and the lack of clinical validation regarding measurement precision. Here, a wearable piezoelectric blood-pressure sensor (WPBPS) is reported, which achieves a high normalized sensitivity (0.062 kPa-1 ), and fast response time (23 ms) for CNAP monitoring. The transfer function of a linear regression model is designed, offering a simple solution to convert the flexible piezoelectric sensor signals into BP values. In order to verify the measurement accuracy of WPBPS, clinical trials are performed on 35 subjects aged from 20 to 80 s after screening. The mean difference between the WPBPS and a commercial sphygmomanometer of 175 BP data pairs is -0.89 ± 6.19 and -0.32 ± 5.28 mmHg for systolic blood pressure (SBP) and diastolic blood pressure (DBP), respectively. By building a WPBPS-embedded wristwatch, the potentially promising use of a convenient, portable, continuous BP monitoring system for cardiovascular disease diagnosis is demonstrated.
More Related Videos
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Blood Pressure
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.