An aptamer-based sandwich assay for detection of alpha-defensin human neutrophil protein 1 on a microfluidic platform
Rishabh Gandotra1, To-Wen Chen2, Feng-Chih Kuo3
1Institute of NanoEngineering and Microsystems, National Tsing Hua University, Hsinchu, Taiwan.
Abstract:
The diagnosis of periprosthetic joint infection (PJI) remains a labor-intensive and challenging issue, with life-threatening complications associated with misdiagnoses. Superior diagnostic approaches are therefore urgently needed, and synovial biomarkers are gaining substantial attention in this capacity. A new aptamer-based sandwich assay was developed where the aptamer probes specific to one such biomarker, alpha-defensin human neutrophil protein 1 (HNP 1), was integrated herein into a new microfluidic platform. The magnetic beads coated with the primary aptamer probe were able to bind the target protein with high affinity and high specificity in synovial fluid and a fluorescent-labelled secondary aptamer were further used to quantify HNP 1 in a sandwich approach. Up to four clinical samples with low volume (∼50 μL each) in a much faster assay including detection within <60 min with 100% accuracy (with totally 13 clinical samples without the need of sample pretreatment) through the use of the aptamer-based sandwich assay were automatically detected on a single chip. The wide dynamic range of this compact device, 0.5-100 mg/L, highlights its utility for future PJI diagnostics in the clinic.