Determination of an air kerma-rate correction factor for the S7600 Xoft AxxentⓇ source model
Autumn E Walter1, Larry A DeWerd2
1Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin, Madison, WI; Department of Human Oncology, School of Medicine and Public Health, University of Wisconsin, Madison, WI.
Purpose:
The purpose of this work was to provide guidance for the lack of an air-kerma rate standard for the S7600 Xoft Axxent® source by providing a correction factor to apply to the National Institute of Standards and Technology (NIST) traceable S7500 well chamber (WC) calibration coefficient before the development of an S7600 standard at NIST.
Methods And Materials:
The Attix free air chamber (FAC) at the University of Wisconsin Medical Radiation Research Center was used to measure the air-kerma rate at 50 cm for six S7500 and six S7600 sources. These same sources were then measured using five standard imaging HDR1000+ WCs. The measurements made with the FAC were used to calculate source-specific WC calibration coefficients for the S7500 and S7600 source. These results were compared to the NIST traceable calibration coefficients for the S7500 source. The average results for each WC were then averaged together, and a ratio of the S7600 to S7500 WC calibration coefficients was determined.
Results:
The average S7600 air-kerma rate measurement with the FAC was 7% lower than the average air-kerma rate measurements of the S7500 source. On average, the S7500 determined WC calibration coefficients agreed within ±1% of the NIST traceable S7500 values. The S7600 WC calibration coefficients were up to 16% less than the NIST traceable S7500 values. The final correction factor determined to be applied to the NIST traceable S7500 value was 0.8415 with an associated uncertainty of ±8.1% at k = 2.
Conclusions:
This work provides a suggested correction factor for the S7600 Xoft Axxent source such that the sources can be accurately implemented in the clinical setting.