Error identification and compensation regarding the kinematic parameter of the MD-PEF for tibial deformity correction
Guotong Li1, Xiaoqiang Tang2, Jianfeng Li3
1State Key Laboratory of Tribology, Department of Mechanical Engineering, Tsinghua University, Beijing, 100084, China; Beijing Key Lab of Precision/Ultra-Precision Manufacturing Equipment and Control, Tsinghua University, Beijing, 100084, China; Beijing Key Laboratory of Advanced Manufacturing Technology, Faculty of Materials and Manufacturing, Beijing University of Technology, Beijing, 100124, China.
Abstract:
The correction accuracy of an external fixator is crucial to the treatment outcome of deformity correction and patient safety. In this study, the mapping model is established between the pose error and kinematic parameter error of the motor-driven parallel external fixator (MD-PEF). Subsequently, the kinematic parameter identification and error compensation algorithm of the external fixator is established based on the least squares method. An experimental platform based on the developed MD-PEF and Vicon motion capture system is constructed for kinematic calibration experiments. Experimental results show that the correction accuracy of the MD-PEF after calibration is as follows: translation accuracy dE1 = 0.36 mm, axial translation accuracy dE2 = 0.25 mm, angulation accuracy dE3 = 0.27°, and rotation accuracy dE4 = 0.2°. The accuracy detection experiment verifies the kinematic calibration results, which further validates the feasibility and reliability of the error identification and compensation algorithm constructed by the least squares method. The calibration approach used in this work also provides an effective way to improve the accuracy of other medical robots.
Related Concept Videos
Kinematic Equations: Problem Solving
Time-Domain Interpretation of PD Control
Consider the example of control of motor torque. Initially, a positive...
Deformation of Member under Multiple Loadings
In the case of a member with a variable cross-section, the strain is not constant but depends on the position. The deformation of an...