Related Experiment Video
Updated: Aug 5, 2025

Application of Optical Coherence Tomography to a Mouse Model of Retinopathy
Published on: January 12, 2022
Phase-Decorrelation Optical Coherence Tomography Measurement of Cold-Induced Nuclear Cataract
Brecken J Blackburn1, Matthew T McPheeters1, Michael W Jenkins1,2
1Department of Biomedical Engineering, Case Western Reserve University, Cleveland, OH, USA.
Purpose:
The purpose of this work is to determine the sensitivity of phase-decorrelation optical coherence tomography (OCT) to protein aggregation associated with cataracts in the ocular lens, as compared to OCT signal intensity.
Methods:
Six fresh porcine globes were held at 4°C until cold cataracts developed. As the globes were re-warmed to ambient temperature, reversing the cold cataract, each lens was imaged repeatedly using a conventional OCT system. Throughout each experiment, the internal temperature of the globe was recorded using a needle-mounted thermocouple. OCT scans were acquired, their temporal fluctuations were analyzed, and the rates of decorrelation were spatially mapped. Both decorrelation and intensity were evaluated as a function of recorded temperature.
Results:
Both signal decorrelation and intensity were found to change with lens temperature, a surrogate of protein aggregation. However, the relationship between signal intensity and temperature was not consistent across different samples. In contrast, the relationship between decorrelation and temperature was found to be consistent across samples.
Conclusions:
In this study, signal decorrelation was shown to be a more repeatable metric for quantification of crystallin protein aggregation in the ocular lens than OCT intensity-based metrics. Thus, OCT signal decorrelation measurements could enable more detailed and sensitive study of methods to prevent cataract formation.
Translational Relevance:
This dynamic light scattering-based approach to early cataract assessment can be implemented on existing clinical OCT systems without hardware additions, so it could quickly become part of a clinical study workflow or an indication for use for a pharmaceutical cataract intervention.
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Imaging Studies III: Computed Tomography