Summary
A new HPLC and TLC method ensures quality control for [177Lu]Lu-PSMA I&T, a prostate cancer therapy. This validated approach accurately measures radiochemical purity and identifies impurities, improving treatment reliability.
Area of Science:
- Nuclear Medicine
- Radiopharmaceutical Chemistry
- Analytical Chemistry
Background:
- [177Lu]Lu-PSMA I&T (zadavotide guraxetan) is an effective targeted radionuclide therapy for advanced prostate cancer.
- Existing radiochemical purity assays for [177Lu]Lu-PSMA I&T face limitations with standard HPLC methods.
- Trifluoroacetic acid (TFA) in HPLC gradients can cause sample retention and peak tailing.
Purpose of the Study:
- To validate a robust HPLC and TLC method for the quality control of [177Lu]Lu-PSMA I&T.
- To ensure accurate determination of radiochemical purity, identity, and impurity levels.
- To identify the primary radiochemical impurity and assess stability.
Main Methods:
- High-Performance Liquid Chromatography (HPLC) using a Phosphate buffer/Acetonitrile gradient.
- Thin-Layer Chromatography (TLC) with 0.1N Citrate buffer (pH 5).
- Mass spectrometry for impurity identification.
Main Results:
- The validated HPLC method demonstrated accuracy, specificity, robustness, linearity, and adequate limit of quantification (LOQ).
- HPLC analysis yielded symmetrical peaks and quantitative recovery, with radiochemical purity consistently exceeding 95%.
- Stability studies revealed degradation due to radiolysis, mitigated by ascorbic acid, dilution, and cold storage; the main impurity was the de-iodinated form.
Conclusions:
- The combined HPLC and TLC method offers a reliable analytical tool for [177Lu]Lu-PSMA I&T quality control.
- This validated method addresses limitations of previous assays, enhancing therapeutic agent reliability.
- The findings support consistent and safe application of [177Lu]Lu-PSMA I&T in clinical settings.