Related Experiment Video
Updated: Aug 5, 2025

Semiconductor Sequencing for Preimplantation Genetic Testing for Aneuploidy
Published on: August 25, 2019
Non-Invasive prenatal testing with rolling circle amplification: Real-world clinical experience in a non-molecular
Matthew L Saidel1, Uma Ananth2, Donna Rose3
1Women's Health USA and Medical Director, Women's Health Connecticut, Rocky Hill, Connecticut, USA.
Background:
Non-invasive prenatal testing (NIPT) using cell-free DNA (cfDNA) circulating in maternal blood provides a sensitive and specific screening technique for common fetal aneuploidies, but the high cost and workflow complexity of conventional methodologies limit its widespread implementation. A unique rolling circle amplification methodology reduces cost and complexity, providing a promising alternative for increased global accessibility as a first-tier test.
Methods:
In this clinical study, 8160 pregnant women were screened on the Vanadis system for trisomies 13, 18, and 21, and positive results were compared to clinical outcomes where available.
Results:
The Vanadis system yielded a 0.07% no-call rate, a 98% overall sensitivity, and a specificity of over 99% based on available outcomes.
Conclusion:
The Vanadis system provided a sensitive, specific, and cost-effective cfDNA assay for trisomies 13, 18, and 21, with good performance characteristics and low no-call rate, and it eliminated the need for either next-generation sequencing or polymerase chain reaction amplification.