Dejian Li1, Hanru Ren1, Xu Zhang1
1Department of Orthopedics, Shanghai Pudong Hospital, Fudan University Pudong Medical Center, Shanghai 201300, China.
Objective:
This study aims to analyze the biomechanical characteristics of tile B2 pelvic fractures using finite element analysis when the superior ramus of the pubis was fixed by a plate or hollow screws in standing and sitting positions, respectively.
Methods:
A three-dimensional digital model of the tile B2 pelvic fracture was obtained by CT scanning the patient. The main ligament structure was then reconstructed based on the anatomical characteristics to create a finite element model of the tile B2 pelvic fracture. The posterior pelvic ring was fixed by sacroiliac joint screws, while the anterior ring injury of the superior ramus of the pubis was fixed by plates and hollow compression screws, respectively. The degrees of freedom of the bilateral acetabulum or two sides of the ischial tuberosity were constrained in the two models. A vertical load of 600 N was applied to the upper surface of the sacrum to measure the displacement and stress distribution of the pelvis in the standing and sitting positions.
Results:
The displacement distribution of both the healthy and the affected side of the pelvis was relatively uniform in both the plate group and the hollow screw group according to the finite element simulation results. The maximum displacement value in the sitting position was greater than the standing position, and the maximum displacement value of the hollow screw fixation was greater than that of the plate fixation. In the four groups of fixation models, the maximum displacement value of the pelvis in the hollow screw sitting position group was 1616.80 × 10-3 mm, which was greater than that of the other three groups, and in this group the total displacement value of the hollow screw in the anterior ring was 556.31 × 10-3 mm. The stress distribution of the pelvis in the various models was similar in the four groups of models, in which the maximum stress of the pelvis in the hollow screw sitting position group was the largest, which was 201.33 MPa, while the maximum stress in the standing position was 149.85 MPa greater than that in the sitting position of the hollow screw fixation.
Conclusion:
The anterior ring of patients with Tile B2 pelvic fractures fixed with hollow screws or plates in both standing and sitting positions can achieve satisfactory biomechanical results with significant safety margins for plates and screws.
More Related Videos
15:11Surgical Fixation of Sternal Fractures: Preoperative Planning and a Safe Surgical Technique Using Locked Titanium Plates and Depth Limited Drilling
Published on: January 5, 2015
06:38Establishment of a Segmental Femoral Critical-size Defect Model in Mice Stabilized by Plate Osteosynthesis
Published on: October 12, 2016
Related Concept Videos
Stresses under Combined Loadings
The process begins by slicing the tube at critical points and analyzing the internal forces and stress components at these sections, focusing on the centroid. Normal stresses, generated by axial forces and bending moments, are either compressive or tensile and vary across the section from...
Shear on the Horizontal Face of a Beam Element
Plastic Deformations of Members with a Single Plane of Symmetry
Castigliano's Theorem: Problem Solving
Saint-Venant's Principle