Dual-energy CT kidney stone characterization-can diagnostic accuracy be achieved at low radiation dose?
André Euler1, Sara Wullschleger2, Thomas Sartoretti2
1Diagnostic and Interventional Radiology, University Hospital Zurich, University of Zurich, Rämistrasse 100, 8091, Zurich, Switzerland. and.euler@gmail.com.
Objectives:
To assess the accuracy of low-dose dual-energy computed tomography (DECT) to differentiate uric acid from non-uric acid kidney stones in two generations of dual-source DECT with stone composition analysis as the reference standard.
Methods:
Patients who received a low-dose unenhanced DECT for the detection or follow-up of urolithiasis and stone extraction with stone composition analysis between January 2020 and January 2022 were retrospectively included. Collected stones were characterized using X-ray diffraction. Size, volume, CT attenuation, and stone characterization were assessed using DECT post-processing software. Characterization as uric acid or non-uric acid stones was compared to stone composition analysis as the reference standard. Sensitivity, specificity, and accuracy of stone classification were computed. Dose length product (DLP) and effective dose served as radiation dose estimates.
Results:
A total of 227 stones in 203 patients were analyzed. Stone composition analysis identified 15 uric acid and 212 non-uric acid stones. Mean size and volume were 4.7 mm × 2.8 mm and 114 mm3, respectively. CT attenuation of uric acid stones was significantly lower as compared to non-uric acid stones (p < 0.001). Two hundred twenty-five of 227 kidney stones were correctly classified by DECT. Pooled sensitivity, specificity, and accuracy were 1.0 (95%CI: 0.97, 1.00), 0.93 (95%CI: 0.68, 1.00), and 0.99 (95%CI: 0.97, 1.00), respectively. Eighty-two of 84 stones with a diameter of ≤ 3 mm were correctly classified. Mean DLP was 162 ± 57 mGy*cm and effective dose was 2.43 ± 0.86 mSv.
Conclusions:
Low-dose dual-source DECT demonstrated high accuracy to discriminate uric acid from non-uric acid stones even at small stone sizes.
Key Points:
• Two hundred twenty-five of 227 stones were correctly classified as uric acid vs. non-uric acid stones by low-dose dual-energy CT with stone composition analysis as the reference standard. • Pooled sensitivity, specificity, and accuracy for stone characterization were 1.0, 0.93, and 0.99, respectively. • Low-dose dual-energy CT for stone characterization was feasible in the majority of small stones < 3 mm.
More Related Videos
07:013D Imaging of Soft-Tissue Samples using an X-ray Specific Staining Method and Nanoscopic Computed Tomography
Published on: October 24, 2019
08:41Novel In Vivo Micro-Computed Tomography Imaging Techniques for Assessing the Progression of Non-Alcoholic Fatty Liver Disease
Published on: March 24, 2023
Related Concept Videos
Imaging Studies III: Computed Tomography
Radiological Investigation I: X-ray and CT
Imaging Studies II: Ultrasonography
Imaging Studies I: Kidney, Ureter, and Bladder Studies
Imaging Studies for Cardiovascular System V: CT
Urinary Tract Calculi III: Medical Management